24
Thank you for purchasing an
MiniFlex
®
MA110 or MA130
flexible current sensor.
For best results from your instrument:
read
these operating instructions carefully,
comply
with the precautions for use.
WARNING, risk of DANGER! The operator must refer to these instructions whenever this
danger symbol appears.
WARNING, risk of electric shock. The voltage applied to parts marked with this symbol may
be hazardous.
Equipment protected by double insulation.
Earth.
Application or withdrawal authorized on conductors carrying dangerous voltages. Type B current
sensor as per IEC 61010‑2‑032.
Battery.
Direction of the current.
The product is declared recyclable following an analysis of the life cycle in accordance with
standard ISO14040.
Chauvin Arnoux has adopted an Eco‑Design approach in order to design this appliance. Analysis
of the complete lifecycle has enabled us to control and optimize the effects of the product on
the environment. In particular this appliance exceeds regulation requirements with respect to
recycling and reuse.
The CE marking indicates conformity with European directives, in particular LVD and EMC.
The rubbish bin with a line through it indicates that, in the European Union, the product must
undergo selective disposal in compliance with Directive WEEE 2002/96/EC.
Definition of measurement categories:
Measurement category IV corresponds to measurements taken at the source of low‑voltage installations.
Example: power feeders, counters and protection devices.
Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit‑breakers, machines or fixed industrial devices
Measurement category II corresponds to measurements taken on circuits directly connected to low-
voltage installations.
Example: power supply to electro‑domestic devices and portable tools.
ENGLISH