Cleaning, Maintenance, Overhaul
18
5
Cleaning, Maintenance, Overhaul
5.1
Cleaning and overhaul
Regular cleaning ensures perfect hygiene and operation of the Intelect RPW Lite.
CAUTION!
Electrical hazard!
Disconnect the device from the mains before starting any cleaning
and overhauling work!
• Unplug the mains plug !
The frequency of complete exterior cleaning depends on the frequency of use and
what the device is used for.
All parts which come into contact with the patient must be cleaned
before and
after
each treatment.
• Wipe down the device parts with a damp cloth.
• For cleaning, use a lukewarm, dilute solution of non-vegetable soapy water.
ATTENTION!
It is essential that no fluid be permitted to penetrate either the device or its
tubing.
• Keep the ventilation slits clear.
Cleaning the handpiece
The handpiece, in particular the pulse transmitter, must be thoroughly cleaned and
disinfected after each therapy session.
• Clean the coupling gel off the handpiece using a cleaning agent that is suitable
for surface cleaning.
• Disinfect the handpiece with an alcohol-based disinfectant that is suitable for
surface cleaning (follow the manufacturer's instructions).