57
Triton® Traction Unit
7 CALIBRATION
6.
Press the Clinical Resources button. See
Figure
7.3.
7.
Simultaneously press the START, STOP, and
Clinical Resources buttons. See Figure 7.4.
8.
Press "Zero the Load Cell" on Touch Screen.
See
Figure
7.5.
9.
When the screen message appears, "Load
Cell has been zero calibrated!", touch the
Touch Screen anywhere to continue.
7.1 UNIT CALIBRATION (continued)
FIGURE 7.3
FIGURE 7.4
FIGURE 7.5
Summary of Contents for TRITON 2841
Page 15: ...13 Triton Traction Unit NOTES ...
Page 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Page 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Page 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Page 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Page 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Page 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Page 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Page 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Page 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Page 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Page 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Page 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Page 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Page 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...