56
Triton® Traction Unit
7 CALIBRATION
A. General Information
1.
The Calibration and Burn-In Procedures in
this section must be performed annually
and when field service requires removal and
replacement of certain components.
2.
The Calibration Procedure requires use of
US lb Weight Measurements. The software
is fixed and will not perform conversions of
Metric Weight Measurements for this critical
procedure. Therefore, when constructing
and equipping the Calibration Fixture, it is
required that the Electronic Load Cell
Components be capable of expressing
US lbs.
3.
The fabrication of a Calibration Fixture is
required for these units.
B.
Equipment
Required
Calibration Fixture with calibrated Strain Meter
C. Unit Calibration Procedure
1.
Place unit on Calibration Fixture and secure
in position by tightening the unit Lock
Knob. See Figure 7.1.
2.
Connect unit to an approved power source.
3.
Turn unit On.
4.
Release Traction Cord and connect
Accessory Clip to the Strain Meter. Leave
slack in Cord. See Figure 7.2.
5.
Zero Strain Meter LCD.
7.1 UNIT CALIBRATION
FIGURE 7.1
FIGURE 7.2
Certain removal and replacement procedures require that complete Unit Calibration and Burn In be performed
prior to placing the unit back into service.
Failure to perform either of these procedures after certain component or assembly replacement will void
warranty and may result in injury to patients and severe equipment damage.
When performing Unit Calibration Procedure, the
Calibration Fixture Load Cell must be set to US lbs.
ACCESSORY CLIP TO
LOAD CELL
STRAIN METER
SLACK IN
CORD
Summary of Contents for TRITON 2841
Page 15: ...13 Triton Traction Unit NOTES ...
Page 72: ...Triton Traction Unit 70 9 SCHEMATICS Control Board 47801 1 of 3 ...
Page 73: ...Triton Traction Unit 71 9 SCHEMATICS Control Board 47801 2 of 3 ...
Page 74: ...Triton Traction Unit 72 9 SCHEMATICS Control Board 47801 3 of 3 ...
Page 75: ...Triton Traction Unit 73 9 SCHEMATICS Motor Control Board 47803 1 of 2 ...
Page 76: ...Triton Traction Unit 74 9 SCHEMATICS Motor Control Board 47803 2 of 2 ...
Page 77: ...Triton Traction Unit 75 9 SCHEMATICS Patient Interrupt Switch 48056 1 of 1 ...
Page 78: ...Triton Traction Unit 76 9 SCHEMATICS Serial Interface 47022 1 of 1 ...
Page 79: ...Triton Traction Unit 77 9 SCHEMATICS Smart Card 47807 1 of 1 ...
Page 80: ...Triton Traction Unit 78 9 SCHEMATICS sEMG Module 48123 1 of 1 ...
Page 81: ...Triton Traction Unit 79 9 SCHEMATICS Block Diagram 1 of 1 ...
Page 82: ...Triton Traction Unit 80 10 SPECIAL FIXTURES Calibration Fixture 1 of 4 ...
Page 83: ...Triton Traction Unit 81 10 SPECIAL FIXTURES Calibration Fixture 2 of 4 ...
Page 84: ...Triton Traction Unit 82 10 SPECIAL FIXTURES Calibration Fixture 3 of 4 ...
Page 85: ...Triton Traction Unit 83 10 SPECIAL FIXTURES Calibration Fixture 4 of 4 ...