4
INTELECT LEGEND COMBO
Precautionary Instructions
1.
CAUTION:
Read, understand and practice the precautionary and operating instructions. Know the limitations and hazards
associated with using any electrical stimulation or ultrasound device. Observe the precautionary and operational
decals placed on the unit.
2.
CAUTION:
Do not operate the Intelect
®
Legend Combo when connected to any unit other than Chattanooga Group devices. Do
not operate the unit in an environment of short-wave diathermy use.
3.
WARNING:
Federal law restricts this device to sale by, or on the order of, a physician or licensed practitioner. This device should
be used only under the continued supervision of a physician or licensed practitioner.
4.
CAUTION:
This unit generates, uses and can radiate radio frequency energy, and if not installed and used in accordance with the
instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. Harmful interference to other devices can be determined by
turning this unit on and off or trying to correct the interference using one or more of the following: Reorient or
relocate the receiving device, increase the separation between the equipment, connect the unit to an outlet on a
different circuit from that to which the other device(s) are connected and/or consult the factory field service
technician for help.
5.
CAUTION:
The Ultrasound generator should be routinely checked before each use to determine that all controls
function normally; especially that the intensity control does properly adjust the intensity of the ultrasonic power
output in a stable manner. Also, determine that the treatment time control does actually terminate ultrasonic power
output when the timer reaches zero.
6.
CAUTION:
Use of controls or adjustments or performance of procedures other than those specified herein may result in
hazardous exposure to ultrasonic energy.
7.
CAUTION:
This unit is not designed to prevent the ingress of water or liquids. Ingress of water or liquids could cause
malfunction of internal components of the system and therefore create a risk of injury to the patient.
8.
CAUTION:
DO NOT use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel as
damage may result.
9.
WARNING:
Equipment not suitable for use in the presence of flammable anesthetics mixture with air, oxygen, or nitrous oxide.
The warning symbol for this hazard is prominently displayed on the cabinet.