background image

1

2

1.4.5 Stimulation should not be applied over swollen, infected, or inflamed areas or skin eruptions, e.g., phlebitis,

thrombophlebitis, varicose veins, etc.

1.4.6 Stimulation should not be applied over, or in proximity to, cancerous lesions.
1.4.7 For external use only.
1.4.8 Do not use IFC on the eye area.
1.4.9 This device should be used only under the continued supervision of a licensed medical practitioner.
1.4.10 Safety of IFC devices for use during pregnancy or delivery has not been established.
1.4.11 Electronic equipment such as ECG monitors and ECG alarms may not operate properly when IFC is in use.
1.4.12 Apply the electrodes to clean, dry, and unbroken skin only.
1.4.13 This device should not be used while driving, operating machinery, or during any activity in which

involuntary muscle contractions may put the user at undue risk of injury.

1.4.14 This device should be kept out of the reach of children.
1.4.15 Keep electrodes separate during treatment. Electrodes in contact with each other could result in improper

stimulation or skin burns.

1.5 Precautions:

1.5.1 Caution should be used for patients with suspected or diagnosed heart problems.
1.5.2 Caution should be used for patients with suspected or diagnosed epilepsy.

1.1 General information:
This interferential (IFC) device is a lightweight and portable medical device which can help to reduce pain and
discomfort. It utilizes the low electric-current to stimulate muscle nerves to achieve the symptomatic relief of
chronic intractable pain, post-traumatic pain, and post-surgical pain.

1.2 Cautions
Federal law (USA) restricts this device to sale by or on the order of practitioners licensed by the State in which they
practice to use or order the use of the device.

1.3 Indications for use:
This device is used in symptomatic relief of chronic intractable pain, post-traumatic and post-surgical pain. 

1.4 Warnings: 

1.4.1 The long-term effects of chronic electrical stimulation are unknown.
1.4.2 Stimulation should not be applied over the carotid sinus nerves, particularly in patients with a known

sensitivity to the carotid sinus reflex.

1.4.3 Stimulation should not be applied over the neck or mouth.  Severe spasm of the laryngeal and pharyngeal

muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in
breathing.

1.4.4 Stimulation should not be applied transthoracically in that the introduction of electrical current into the

heart may cause cardiac arrhythmias.

1. INTRODUCTION

Summary of Contents for Intelect IFC

Page 1: ...C Intelect TENS Intelect TENS Intelect TENS 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical NMES Intelect NMES Intelect 77393A 77395A 77684C 77392A 77391A 77394A ...

Page 2: ... 7 ACCESSORIES 16 8 TROUBLESHOOTING 17 9 WARRANTY 18 Chapter Page 1 INTRODUCTION 1 1 General information 1 1 2 Cautions 1 1 3 Indications for use 1 1 4 Warnings 1 2 1 5 Precautions 2 3 1 6 Adverse Reactions 3 2 PRODUCT DESCRIPTIONS 4 6 3 STIMULATION MODES 7 8 4 INSTRUCTIONS FOR USE 4 1 Check Battery 9 4 2 Connect electrodes to lead wires 9 4 3 Connect lead wires to unit 10 4 4 Place electrodes on ...

Page 3: ... 7 ACCESSORIES 16 8 TROUBLESHOOTING 17 9 WARRANTY 18 Chapter Page 1 INTRODUCTION 1 1 General information 1 1 2 Cautions 1 1 3 Indications for use 1 1 4 Warnings 1 2 1 5 Precautions 2 3 1 6 Adverse Reactions 3 2 PRODUCT DESCRIPTIONS 4 6 3 STIMULATION MODES 7 8 4 INSTRUCTIONS FOR USE 4 1 Check Battery 9 4 2 Connect electrodes to lead wires 9 4 3 Connect lead wires to unit 10 4 4 Place electrodes on ...

Page 4: ...rt problems 1 5 2 Caution should be used for patients with suspected or diagnosed epilepsy 1 1 General information This interferential IFC device is a lightweight and portable medical device which can help to reduce pain and discomfort It utilizes the low electric current to stimulate muscle nerves to achieve the symptomatic relief of chronic intractable pain post traumatic pain and post surgical ...

Page 5: ...rt problems 1 5 2 Caution should be used for patients with suspected or diagnosed epilepsy 1 1 General information This interferential IFC device is a lightweight and portable medical device which can help to reduce pain and discomfort It utilizes the low electric current to stimulate muscle nerves to achieve the symptomatic relief of chronic intractable pain post traumatic pain and post surgical ...

Page 6: ...s recommended for use by the manufacturer 1 5 7 Isolated cases of skin irritation may occur at the site of the electrode placement following long term application 1 5 8 Effectiveness is highly dependent upon patient selection by a person qualified in the management of pain patients 1 5 9 If the stimulation levels are uncomfortable or become uncomfortable reduce the stimulation amplitude to a comfo...

Page 7: ...s recommended for use by the manufacturer 1 5 7 Isolated cases of skin irritation may occur at the site of the electrode placement following long term application 1 5 8 Effectiveness is highly dependent upon patient selection by a person qualified in the management of pain patients 1 5 9 If the stimulation levels are uncomfortable or become uncomfortable reduce the stimulation amplitude to a comfo...

Page 8: ...at the right top of the unit regulates the amplitude or intensity of the stimulation and is the ON OFF CONTROL The power indicator will light up with green color when the unit is working Caution If the stimulation levels are uncomfortable or become uncomfortable reduce the stimulation intensity to a comfortable level and contact your physician if problems persist Rate Control Adjust Frequency from...

Page 9: ...at the right top of the unit regulates the amplitude or intensity of the stimulation and is the ON OFF CONTROL The power indicator will light up with green color when the unit is working Caution If the stimulation levels are uncomfortable or become uncomfortable reduce the stimulation intensity to a comfortable level and contact your physician if problems persist Rate Control Adjust Frequency from...

Page 10: ...switch to adjust the mode Be sure that when adjusting these stimulation modes the intensity Amplitude output controls are set to the minimum output positions 3 1 C Continuous In the C Continue mode there is no change in the pulse rate When set at the other modes the interference frequency changes over time For example when the frequency control was set at 100Hz the interference frequency would shi...

Page 11: ...switch to adjust the mode Be sure that when adjusting these stimulation modes the intensity Amplitude output controls are set to the minimum output positions 3 1 C Continuous In the C Continue mode there is no change in the pulse rate When set at the other modes the interference frequency changes over time For example when the frequency control was set at 100Hz the interference frequency would shi...

Page 12: ... or rechargeable battery into the battery compartment Make sure that you are installing the battery properly The battery is inserted in the casing on the foot of the stimulator unit BE SURE TO MATCH THE POSITIVE AND NEGATIVE ENDS OF THE BATTERY TO THE MARKINGS IN THE BATTERY COMPARTMENT OF UNIT CONNECTING THE STIMULATOR 4 2 Connect electrodes to lead wires Insert the lead wire connector into elect...

Page 13: ... or rechargeable battery into the battery compartment Make sure that you are installing the battery properly The battery is inserted in the casing on the foot of the stimulator unit BE SURE TO MATCH THE POSITIVE AND NEGATIVE ENDS OF THE BATTERY TO THE MARKINGS IN THE BATTERY COMPARTMENT OF UNIT CONNECTING THE STIMULATOR 4 2 Connect electrodes to lead wires Insert the lead wire connector into elect...

Page 14: ...essional 4 4 Place electrodes on skin Apply electrodes to the exact site indicated by your physician following the instruction included with the electrodes labeling Before applying electrodes be sure the skin surface over which electrodes are placed is thoroughly cleaned and dried Make sure the electrodes are placed firmly to skin and make good contact between the skin and the electrodes Place the...

Page 15: ...essional 4 4 Place electrodes on skin Apply electrodes to the exact site indicated by your physician following the instruction included with the electrodes labeling Before applying electrodes be sure the skin surface over which electrodes are placed is thoroughly cleaned and dried Make sure the electrodes are placed firmly to skin and make good contact between the skin and the electrodes Place the...

Page 16: ...s with C E mark or which are legally marketed in the US under 510 K procedure 4 13 Care of Electrode cords Clean the electrode cords by wiping them with damp cloth Coating them lightly with talcum powder will reduce tangles and prolong the life CARE AND MAINTENANCE 4 10 Portability Your unit is portable and may be clipped to a belt shirt pocket bra or other clothing 4 11 Battery To replace the bat...

Page 17: ...s with C E mark or which are legally marketed in the US under 510 K procedure 4 13 Care of Electrode cords Clean the electrode cords by wiping them with damp cloth Coating them lightly with talcum powder will reduce tangles and prolong the life CARE AND MAINTENANCE 4 10 Portability Your unit is portable and may be clipped to a belt shirt pocket bra or other clothing 4 11 Battery To replace the bat...

Page 18: ...ery Size D x W x H 1 3 x 2 5 x 3 7 33 mm x 64 mm x 94 mm Weight including battery 4 7 oz 134 g Low Battery Indication Yes Safety Standard EN 60601 1 EN 60601 1 2 IEC 60601 2 10 Operation Ambient Temp Range 50 95ºF 10 35ºC Operation Ambient humidity Range 20 90 RH Storage Transportation Temp Range 32 158ºF 0 70ºC Storage Transportation Humidity Range 20 90 RH All values have 10 tolerance 16 15 6 SP...

Page 19: ...ery Size D x W x H 1 3 x 2 5 x 3 7 33 mm x 64 mm x 94 mm Weight including battery 4 7 oz 134 g Low Battery Indication Yes Safety Standard EN 60601 1 EN 60601 1 2 IEC 60601 2 10 Operation Ambient Temp Range 50 95ºF 10 35ºC Operation Ambient humidity Range 20 90 RH Storage Transportation Temp Range 32 158ºF 0 70ºC Storage Transportation Humidity Range 20 90 RH All values have 10 tolerance 16 15 6 SP...

Page 20: ... incidental or consequential damages so the above limitation or exclusion may not apply to you 8 TROUBLESHOOTING 18 17 The power indicator lights up but unit does not function properly On and Battery Light are dim None of indicators light up 1 Check all control settings Are they set to values prescribed by your medical professional 2 Are electrodes in proper position 3 Check lead wires Be sure all...

Page 21: ... incidental or consequential damages so the above limitation or exclusion may not apply to you 8 TROUBLESHOOTING 18 17 The power indicator lights up but unit does not function properly On and Battery Light are dim None of indicators light up 1 Check all control settings Are they set to values prescribed by your medical professional 2 Are electrodes in proper position 3 Check lead wires Be sure all...

Page 22: ...C Intelect TENS Intelect TENS Intelect TENS 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical 2003 Encore Medical NMES Intelect NMES Intelect 77393A 77395A 77684C 77392A 77391A 77394A ...

Reviews: