CALIBRATION PROCEDURE
To calibrate the scale, please turn off the scale.
Operation
Display
Turn the scale on using the ON/OFF key.
Press UNIT key for 3 seconds and SETUP
will show shortly. UNIT will appear on the
display.
Press the TARE key repeatedly until the
“CAL ib” symbol is displayed.
Press the HOLD key.
Press the TARE key.
The triangle [
] will display on the top of
left hand side of the display. If not, press
the TARE key.
Press the HOLD key repeatedly until the
“CAL 0” symbol is displayed.