3 SCPI
3.3 Instrument-specific Commands
56
Reset status:
20 dB
Key path:
Front panel
【
Mode Setup
】
—>[Uncertainty Calculator]—>[DUT Gain]
[ DUT Input Match ]
:CALCulate:UNCertainty:DUT:MATCh:INPut
Function description:
This command is used to set/query the input match of the device under test
without any unit.
Setting format:
:CALCulate:UNCertainty:DUT:MATCh:INPut <Val>
Query format:
:CALCulate:UNCertainty:DUT:MATCh:INPut?
Return value:
Input match value
Parameter description:
<Val> Input match value, range: [-100, 100] dB.
Example:
:CALC:UNC:DUT:MATC:INP 2
:CALC:UNC:DUT:MATC:INP?
Reset status:
1.50
Key path:
Front panel
【
Mode Setup
】
—>[Uncertainty Calculator]—>[DUT Input Match*]
[ DUT Output Match ]
:CALCulate:UNCertainty:DUT:MATCh:OUTPut
Function description:
This command is used to set/query the output match of the device under test
without any unit.
Setting format:
:CALCulate:UNCertainty:DUT:MATCh:OUTPut <Val>
Query format:
:CALCulate:UNCertainty:DUT:MATCh:OUTPut?
Return value:
Input match value
Parameter description:
<Val> Input match value, range: [-100, 100] dB.
Example:
:CALC:UNC:DUT:MATC:OUTP 2
:CALC:UNC:DUT:MATC:OUTP?
Reset status:
1.50
Key path:
Front panel
【
Mode Setup
】
—>[Uncertainty Calculator]—>[DUT Output Match*]
[ DUT Noise Figure ]
:CALCulate:UNCertainty:DUT:NFIGure
Function description:
This command is used to set/query the noise figure of the device under test. The
default unit is dB.
Setting format:
:CALCulate:UNCertainty:DUT:NFIGure <Val>
Query format:
:CALCulate:UNCertainty:DUT:NFIGure?
Return value:
Noise figure
Parameter description: