![cetoni NEM-B124-02 A Hardware Manual Download Page 10](http://html1.mh-extra.com/html/cetoni/nem-b124-02-a/nem-b124-02-a_hardware-manual_2584871010.webp)
10
CETONI Nemesys S Hardware Manual
2.4.4
Safety measures
The safety of the user and a failure-free operation of the CETONI Nemesys S are assured only if original
parts are used. Only original accessories may be used. Warranty claims will not be accepted for damage
due to the use of alien accessories or expendables.
The CETONI Nemesys S has been developed and constructed in such a way as to largely rule out hazards
due to its intended use. Nevertheless, you must observe the following security measures in order to
exclude any remaining hazards:
•
CETONI GmbH points out the responsibilities of the operator for the operation of the devices.
The laws and regulations of the place of installation must be observed while operating the
devices! To ensure a safe work routine, operators and users must assume responsibility for
adhering to regulations.
•
Before operating the device, the user must ensure the functional safety and proper condition of
the device, its accessories and the cables. The device must not be operated if there is any visible
damage.
•
The user must be familiar with the operation of the devices and the software.
•
Cables must be laid in a way that avoids any risk of stumbling.
•
Any moving parts must not be touched whilst the devices are in operation. There is a risk of
crushing!
•
Relieve the pressure in the system before loosening connections.
•
Check the leak tightness of all fluidic connections after connection and at regular intervals.
•
Only use connection material that is specified for the expected pressures.
•
The device is designed and approved to work in fluidic systems, which fall within the scope of
Article 4 Paragraph 3 of the Pressure Equipment Directive 2014/68/EU. This means that the
system may not exceed a maximum volume of 1 liter. With the use of fluids from Group 1
according to Article 13 of the Pressure Equipment Directive 2014/68/EU, the maximum allowable
system pressure is 200 bar. For fluids from Group 2 it is 1000 bar. If different, product-specific
values for the maximum pressure are given in the section "Technical Data", these values must be
complied with. Regarding the maximum operating temperature, the specification from the
section "Technical Data" must be observed.
CETONI GmbH is not liable for consequences that may arise if the user expands the system by
peripheral devices, such that one of the values or both values are exceeded.