Centurion Scientific Pro-Research Instruction Manual Download Page 1

Centuri

o

n

Scientific Limited

LCD

 display

 

centrifuges

PrO-

Research

Ambient 

& Refrigerated 

instruction manual

Summary of Contents for Pro-Research

Page 1: ...Centurion Scientific Limited LCD display centrifuges PrO Research Ambient Refrigerated instruction manual...

Page 2: ...ble and free from resonance The Bench or supporting structure must be suitable for horizontal use of the centrifuge The Centrifuge should Not be exposed to heat or strong sunlight Ultra violet rays ca...

Page 3: ...ft and fit rotor onto the motor shaft M Fit theScrew and springwasher screw down tightwith the supplied tool Checktightness regularly Fixing for micro and fixed angle rotors BRK5401 BRK54XXseries BRK5...

Page 4: ...Tare zero your scales first Weigh a tube note its weight Weigh another and match its weight to within 0 1 gram Place all your required tubes in all 4 buckets Equal amount per bucket Weigh each bucket...

Page 5: ...s Part number 8011 CHECK the rotor screw fixing tightness regularly Micro tube Rotor ALWAYS load your tubes opposite to each other ALWAYS weigh samples to within0 1 Gram For best separations seep page...

Page 6: ...he rotor screw fixing tightness regularly Large fixed angle Rotors ALWAYS load your tubes opposite to each other ALWAYS weigh samples to within 0 1 Gram For best separations see page 4 CHECK The rotor...

Page 7: ...tightness regularly ALWAYS load your tubes opposite to each other ALWAYS Weigh Bucket and all samples within to 1 0 Gram ALWAYS fit ALL 4 Buckets and 4 Adaptors supplied CHECK the rotor screw fixing...

Page 8: ...lanced NEVER run swing out rotors with only two buckets All four buckets must be used at all times Wrong Wrong Tube placements Swing out rotor Always use the same size tubes as in A per bucket NEVER u...

Page 9: ...p Gaskets and replace regularly Part Number 8055 CHECK the rotor screw fixing tightness regularly ALWAYS load your tubes opposite to each other ALWAYS weigh samples to within 0 2 Gram For best separat...

Page 10: ...to make sure that opposite tubes are of the same weight to avoid rotor imbalance If the tubes are filled with the same material they must therefore be filled with the same amount No changes should be...

Page 11: ...ed Display shown after pressing Tick on name settings Display shown after pressing Set Parameters Display shown after pressing Rotor rectangle Display shown when Set Name pressed Display shown when ch...

Page 12: ...pressed Display shown Display shown when RPM pressed Display shown after selecting RPM SELECT and PRESS desired timer type Time 0 9999 minutes RCFt See W2t document for explanation Pulse Timer counts...

Page 13: ...cal rectangle Display shown when Acc pressed Display after selecting Acc Display Display Display shown when Dec pressed Display after selecting Acc and Dec rates then return to set Acceleration and De...

Page 14: ...bol This will run the centrifuge at 20 of the maximum speed of the rotor for 20 minutes a will show in the Rpm section to show the Pre cool mode is in operation It is advisable to let the time run to...

Page 15: ...RESS Start button arrow triangle bottom right If you require early termination Time Hold RCFt PRESS Square bottom right OR in pulse mode RELEASE when counter seconds has reached desired number PRESS r...

Page 16: ...screen Shown below It reads as follows either 0 or 1 may show when this screen first appears follow instructions below to perform diagnostic checks or greater version number of installed firmware 0 or...

Page 17: ...Acceleration Rate Set Deceleration Rate Set Time Set RCFt Not set RCF g Set Stop User name Actual Temperature Refrigerated only Temp Set Refrigerated only Time Set Programmed Rotor Set RCFt Not set R...

Page 18: ...ollows Make sure the Rotor has stopped view via the port in the Lid Once stopped turn off the power via the inlet Switch Pull the Centrifuge slightly over the bench as shown in a You will notice one b...

Page 19: ...point on the Speed Scale B Read the point at which the straightedge intersects the Relative Centrifugal Force Scale C in this case 1000 x gravity Similarly if the desired r c f is known the necessary...

Page 20: ...ore running the unit Keep a 30Cm distance around your Centrifuge Do not site the unit in sunlight Do not use any kind of scourers or abrasive material to clean the Centrifuge and ensure that a pH neut...

Page 21: ...using the unit with an unbalanced rotor can cause distortion and excessive wear Read tube placements section in the previous section Do not use any kind of scourers or abrasive material to clean the b...

Page 22: ...ON SCIENTIFICL I M I T E D C E N T U R I O N S C I E N T I F I C L I M I T E D CERTIFICATE OF 21 250 7 Scientific Limited 62 62 EVL HYLFH ODVVLILFDWLRQ DERUDWRU XLSPHQW HQWXULRQ 6FLHQWLILF WG HQWULIXJ...

Page 23: ...y us are of UNITED KINGDOM origin Signed for and on behalf of Centurion Scientific Ltd Granted 20 RQZDUGV Company Representative The Old Stables Church Farm Stoughton Chichester West Sussex PO18 9JL U...

Page 24: ...solution of centrifugation Whilst our centrifuges come with a confidence giving warranty yearly services are imperative to ensure your warranty is valid this also ensures your machine is working to it...