PRECISION WEIGHING

SCALES

User’s Manual

Summary of Contents for PB-60-1500

Page 1: ...Ref 49 MCEPB60EN02 Rev 02 PB 60 PRECISION WEIGHING SCALES User s Manual...

Page 2: ......

Page 3: ...3 1 GENERAL INSTALLATION 3 3 2 INSTALLING THE PB 60 SERIES 3 4 DESCRIPTION OF KEYBOARD 4 5 OPERATION 5 5 1 MANUAL ZERO SETTING 5 5 2 SETTING THE TARE 5 5 3 SAMPLE WEIGHT 5 5 4 PERCENTAGE WEIGHT 5 5 5...

Page 4: ......

Page 5: ...se including item counting and percentage weight functions The series consists of 4 different models with a capacity of up to 3 000 g They all have stainless steel plates on an ABS base All the keypad...

Page 6: ...c tl t t 2 2 Specifications common to all models Interface RS 232 output optional Stabilisation time Normally 2 seconds Operating temperature 0 C 40 C 32 F 104 F Power supply external 9V 800mA or inte...

Page 7: ...cales or submerge them in water If the scales come into contact with water measurements may become unstable or the scales may not operate correctly Please unplug the scales immediately from the power...

Page 8: ...ng mode Prints results on a PC or printer using the optional RS 232 interface It can also be used to add the value of the accumulative memory if the accumulation function is not automatic When the sca...

Page 9: ...s value will be the gross weight of the packaging plus all the product that was removed the zero indicator will also light up because the platform returns to the same situation as when was pressed 5 3...

Page 10: ...total number pcs Press to return to normal weighing mode 5 6 ACCUMULATION When the scales are set in manual accumulation mode the weight that appears on the display can be saved in the memory by pres...

Page 11: ...ation Never use a damaged battery charger or battery Never short circuit the battery An accidental short circuit may occur if a metallic object coin paper clip ballpoint pen makes a direct connection...

Page 12: ...on the possibility of include a RS 232 serial interface Specifications RS 232 interface for weighing data ASCII code 9600 bauds 8 data bits Non parity Connector 9 pin SUB D Type Pin 2 Tx Pin 3 Rx NOT...

Page 13: ...essing the display will show the PIN again Press and to enter The display will show UNLOAD Remove any weight from the tray and press to set zero The display will then request the first weight for cali...

Page 14: ...ight on the tray when the scales are set to zero Incorrect calibration of scales Damaged load cell Damaged electronic components Err 5 Key error Incorrect scale operation Err 6 The A D counts are not...

Page 15: ...e aforementioned manufacturer declares that the apparatus described herein complies with the requirements contained in Directive 89 336 CEE and 73 23 CEE and where applicable to the following harmonis...