Vertical Configuration
1. Attach the load cell assembly to the Motor Bracket B with an M5x12 SHCS using a 4mm
hex key
2. Ensure the measuring end of the load cell is exposed and attach the Low Force
compression platen directly to the load cell
Finger tighten all components. Beware of exerting excessive off-axis and torquing forces
to avoid damage to the load cell.
3. Attach the M5 Compression Base Riser
– Tall to the baseplate of the UniVert with an
M5x16 SHCS using 4mm hex key. This may be easier to do with the UniVert in
horizontal configuration.
4. A Zero Position Calibration is required after changing grips or platens. See the
System
Calibration and Advanced Tools
section of this manual for more details.