Command Quick Reference
G-3
Microstep-Enabled Firmware Commands
Command
Value
Description
N
<n> = 0 or 1
Sets the microstep mode off or on.
S
<n> = 0..40
Sets speed
V
<n> = 5..6000
Sets end velocity in HS/s
v
<n> = 50..1000
Sets start velocity in HS/s
c
<n> = 50..2700
Sets cutoff velocity in HS/s
L
<n> = 1..20
Sets ramp slope
K
<n> = 0..31 (default 11)
Backlash
k
<n> = 0..80 or 0..640 in microstep mode
Report Commands
Command
Description
Q
Query, Status and Error Bytes
?
Reports absolute plunger position
?1
Reports start velocity
?2
Reports top velocity
?3
Reports cutoff velocity
?4
Reports actual position of plunger
?12
Reports number of backlash steps
?13
Reports status of input #1
?14
Reports status of input #2
?22
Reports current value from fluid sensor
F
Reports buffer status
&
Reports firmware version
#
Reports firmware checksum
Error Codes
Command
Description
Notes
0
Error free condition
1 (01h)
Initialization error
Fatal error. Reinitialize pump before resuming normal operation.
2 (02h)
Invalid command
3 (03h)
Invalid operand
4 (04h)
Invalid command sequence
6 (06h)
EEPROM failure
7 (07h)
Device not initialized
9 (09h)
Plunger overload
Fatal error. Reinitialize pump before resuming normal operation.
10 (0Ah)
Valve overload
Fatal error. Reinitialize pump before resuming normal operation.
11 (0Bh)
Plunger move not allowed
15 (0Fh)
Command overflow
Summary of Contents for XP 3000
Page 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Page 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Page 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Page 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Page 134: ......