ALLARME:
TEMPERATURA ELEVATA
PULI-JET
PULI-JET
PL
US
PL
US
CA
TT
ANI GR
OUP
14
15
11
ENGLISH
27
Extraordinary maintenance
Extraordinary maintenance must be entrusted to a trained
technician in possession of original spare parts:
- pay special attention to danger signals. Use goggles,
gloves and disposable overalls for personal protection;
- check that routine maintenance has been duly carried out
and make sure that Magnolia products are used;
- before any intervention carry out a series of washes with
Puli-Jet plus disinfectant, then wait 60 minutes for a com-
plete disinfectant action. For more safety, it is advisable
to disinfect the evening before the intervention;
•
Every 12 months:
- check maximum detected temperatures and all alarms.
Intervene accordingly;
- check the aspirator noise level (max. 72 dB according to
the regulation 3047 (E);
- remove dust from the circuit AC 80 inside electronic
components using a blast of dry air not exceeding 2 bar
pressure. By means of a 6 bar blast of air clean also
the small holes on the frontal cover of the aspirating
unit
(15)
;
- check the conditions of plastic hoses, in particular
the hoses under pressure connecting the centrifugal
separator and the Hydrocyclone ISO. We suggest to
replace these hoses every 12-18 months.
•
Every 18-24 months:
- check the working conditions of the centrifugal separator
(11)
and re-circulation valve
(14)
.
• Every 10.000 working hours:
- every time a component containing a rubber part (“O” ring,
gasket or seal) is disassembled, replace the rubber part;
- replace the motors bearings;
- where routine maintenance is not carried out properly or
unsuitable products are used, train the staff and inform
the person in charge.