a)

b)

Green LED On

The calibrator is switched on
and is operating normally.

Green LED Flashing

Battery level is low (< 1.75 V)
Replace the batteries.

Upper Blue LED On

The calibrator is set to 114 dB and
the level is stable.

Upper Blue LED Flashing

The level is not stable.

Lower Blue LED On

The calibrator is set to 94 dB and
the level is stable.

Lower Blue LED Flashing

The level is not stable.

calibrate using a level at least 20 dB higher than the background noise level.

Lower Blue LED - 94 dB level indicator

Timeout

1. The calibrator will switch off after 1 minute or 10 seconds after the microphone is 

removed.

2. If the calibrator is unstable (there is no microphone inserted), it will switch off after 

10 seconds.

3. If a longer calibration period is required, when you switch the calibrator 

On

, hold 

the On/Off button until both Blue LEDs come on, and then release the On/Off 
button. The calibration signal will now be continuous regardless of whether the 
calibrator is stable or not.

Operation

Field Accuracy Check (Acoustic Calibration)

Follow the instructions below to perform a field accuracy check (acoustic calibration) 
with the CEL-120/1 or CEL-120/2 immediately before and after measurements are 
made with a sound level meter or a sound measuring system.

Caution – 

Make sure the microphone and the calibrator (plus the coupler when used) 

are correctly aligned during insertion because the microphone could be damaged if 
not inserted squarely, or if inserted at speed or if excessive force is used.

1.

Carefully insert either:-

a)

WS2 (half-inch) microphone up to the stop within the calibrator cavity.

Or

b)

CEL-4726 Coupler up to the stop within the calibrator cavity and then 
carefully insert the WS3 (quarter-inch) microphone up to the stop within 
the cavity of the coupler.

Note that the coupler flange does not fit flush against the calibrator rim.

2.

Press the 

On/Off

 button to switch the calibrator on.

Refer  to  the  measuring  instrument’s  handbook  and  select  a  suitable 

measuring range and frequency weighting for the calibration).

3.

Press the 

Level toggle

 button to toggle between the 114.0 dB and 94.0 dB 

levels (

this feature is available only on the CEL-120/1

).

If necessary, adjust the calibration control of the measuring instrument until it 

The  CEL-120  provides  a  highly  stable  reference  level  across  changes  in  all 
environmental conditions. Corrections for variations in atmospheric pressure are not 
required.

Both calibrators generate a highly stable sine wave excitation in the acoustic cavity, 
with low harmonic distortion. 

These calibrators offer a nominal pressure calibration level of 114.0 dB at a frequency 
of 1 kHz. The CEL-120/1 can also supply a calibration level of 94.0 dB at a frequency 
of 1 kHz.

A complete “Acoustic Calibrator” consists of the following items.

• CEL-120/1 Calibrator (Class 1) or  CEL-120/2 Calibrator (Class 2)

• 016022 (2 off) AAA battery (LR03)

• FG40 Field Guide

Optional 1/2”-to-1/4” Microphone Coupler (CEL-4726)      

Schedule of Parts

Preparation for Use

Ensure the battery polarities are correct. Replace the cover after battery installation.

The calibrator is now ready for use.

LEDs

Green LED - Power indicator

For the most reliable operation with longest life, the use of an alkaline battery is 
recommended. Rechargeable batteries may also be used.

It is recommended to remove the batteries if the calibrator is to be out of service for 
long periods.

Upper Blue LED - 114 dB level indicator

If a microphone is not inserted, or if it is not fully inserted, the Blue LED will continue to 
flash  and  the  calibrator  will  not  be  able  to  maintain  control.  When  this  condition 
persists for 10 seconds, the calibrator automatically switches off

After a stable calibration signal has been achieved, the Blue LED will stop flashing 
and will remain on. 

CEL-120/1 only:

On the 

CEL-120/1 only

, select the required calibration level. 

In noisy environments the 114.0 dB level should be used. As a general rule, try to 

shows the correct indication (114.0 dB or 94.0 dB).

4.

Wait 3 seconds for the output to stabilise.

5.

After use press the 

On/Off 

button to switch the calibrator off.

6.

Remove the microphone and any coupler from the calibrator.

The  calibrator  will  also  automatically  switch  itself  off  after  10  seconds  when  the 
microphone is removed.

Product Care

Compliance

CEL-120 Calibrators are precision instruments and should be treated with care. 

Avoid:

• Physical shocks or drops

• Ingress of dirt, dust and other foreign objects into the sound cavity

• Exposure to water droplets or condensation

The CEL-120/1 and CEL-120/2 Acoustic Calibrators comply with the EMC Directive 
2004/108/EC  of  the  European  Union.  They  have  been  tested  according  to  the 
standard delivery schedule and comply with the following standards:

• EN  61000-4-2:2009  Testing  and  measuring  techniques  –  electrostatic 

discharge immunity tests

• EN 61000-4-3:A1+A2:2010 Electromagnetic field immunity tests

• EN 61000-6-3:2011 Emission standard for residential, commercial and light 

industrial environments

114 dB

94 dB