INSTALLATION
V
ENTILATION
A
SSESSMENT
R
EQUIRED
The ventilation system for the hazardous location the oven will be installed in must prevent buildups of
potentially explosive gas or vapor mixtures in the air. Conditions under which this could occur include:
•
Excessive amounts of solvents or other hazardous materials evaporating from samples
while being loaded into the oven chamber.
•
Residual hazardous materials outgassing from treated product or samples while being
unloaded from the oven after processing.
•
Explosive concentrations arising from spilled hazardous materials.
•
Contents evaporating from hazardous materials containers accidentally left open.
Determining Sufficient Ventilation
This determination must be
based on an evaluation of your site
with its airflow and HVAC
characteristics, the specifics of your baking application and the hazardous materials involved. The
HVAC or other climate control system should meet all applicable hazardous location requirements
for the safety jurisdiction you are located in.
General considerations:
•
Consult with your fire or other safety jurisdiction officials to verify the applicable ventilation
requirements and the local interpretations of those requirements.
•
The airflow or active ventilation provided at the front of the oven must be sufficient to
prevent the buildup of explosive gas or vapor concentration from samples, products, or
solvents being handled by the oven operators and other process technicians.
Neutral Gas Ventilation
•
If you are using gas nitrogen (GN
RR
2
RR
) for purging and pressurizing the oven safety
enclosures or for backfilling, ensure there is sufficient ventilation in the workspace area to
prevent asphyxiation or hypoxia conditions.
•
The ventilation must be sufficient to protect employees in the event of a leak in the GN
RR
2
supply system.
•
The workspace ventilation system must provide a minimum of
6 air changes per
minute.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......