31
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INTRODUCTION
S
ERVICE
R
EQUIREMENTS
For repairs, overhauls, scheduled preventative maintenance, and other service-level procedures.
IEC Ex Certified Service Facility Required
The oven must be serviced by an IEC Ex-certified facility or by IEC Ex-certified technicians from
such a facility deemed competent to work with the explosion protection types used in this oven (Ex
ia, pzc, mc).
•
The oven user (the owner or designated responsible individual) is obligated under IEC
60079-19 4.3.4 to verify that the service facility chosen to work on the oven demonstrates
compliance with IEC 60079-19.
•
Your safety jurisdictions’ health and safety laws or ordinances may also require this
compliance.
Explosion Protection Systems Compromised
•
Failure to have the oven serviced by an IEC Ex competent facility can lead to the oven
explosion protection systems being compromised, endangering your employees and facility.
Loss of Ex Marks and Safety Certification
•
The oven Ex marking and Ex explosion certifications are no longer valid if the oven is
serviced by a non-certified facility. This may void or adversely impact insurance policies
and regulatory compliance standing.
•
An oven that loses its Ex certification must be re-certified by an IEC Ex qualified NRTL or
accredited or notified body.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......