25
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INTRODUCTION
Temperature Class
•
The oven is IECEx 60079 temperature class 2 equipment (T2), including under UL, CSA,
and ATEX Differences.
•
The oven is also UL Class 1 Division 2 temperature class 2 equipment (T2).
These classifications allow a maximum internal surface temperature of no more than
300°C
(572°F
). The maximum target temperature setpoint this oven can be set to
200°C (392°F)
.
Hottest Internal Surface Temperature
Hotspots exceeding the temperature displayed on the oven control panel will form on the walls,
ceiling, and floor of the oven chamber while the oven is heating. The oven is engineered so these
hot spots never exceed the displayed temperature by the following amount:
•
+100°C (+212°F)
Maximum Process Temperature
A safety interval must be maintained between the hottest baking process temperature setpoint and
the ignition point of flammable material in the oven chamber.
•
The highest baking application or process temperature setpoint must stay
at least 125°C
(257°F) lower
than the auto-ignition temperature of the flammable material.
Data Plate Markings
The hazardous location markings on the unit data plate are the ultimate indicators of whether the
oven is compatible with your hazardous location zone and materials type. Always check the data
plate prior to installing the oven in a hazardous location.
Summary of Contents for CVO-5-EX Series
Page 1: ...Flammable Solvent Ovens Installation Operation Manual CVO 5 EX Oven Series...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 Special Conditions Of Use 131...
Page 7: ...7 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 11: ...11 P a g e cascadesciences com Tel 503 847 9047 SAFETY CERTS AND STATEMENTS...
Page 17: ...17 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 22: ...22 P a g e cascadesciences com Tel 503 847 9047...
Page 24: ...24 P a g e cascadesciences com Tel 503 847 9047 These groups include propane and n butane...
Page 29: ...29 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 37: ...37 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION...
Page 39: ...39 P a g e cascadesciences com Tel 503 847 9047...
Page 44: ...44 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 51: ...51 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 54: ...54 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 61: ...61 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 65: ...65 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 67: ...67 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 69: ...69 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 72: ...72 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 95: ...95 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 104: ...104 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 116: ...116 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 123: ...123 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 125: ...125 P a g e cascadesciences com Tel 503 847 9047 Continued next page...
Page 127: ...127 P a g e cascadesciences com Tel 503 847 9047 End of Procedure...
Page 128: ...128 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 130: ...130 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS...
Page 132: ......