24 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
S
HELVING
I
NSTALLATION
CVO-5s
To ensure accurate temperature measurement,
one shelf bottom must be in close proximity to the
oven temperature probe.
This probe extends out from the chamber back wall. Do not place the
shelf in direct contact with the probe. Install the shelves evenly spaced in the oven chamber to
maximize temperature uniformity. Please see the orientation image on page 15
1.
Install the shelf clips in the slots of the shelf standard mounting rails located on the sides of
the chamber interior, 4 clips per shelf.
a.
Squeeze each clip, insert the top tab first, and then the bottom tab using a
rocking motion.
b.
Set the shelves on the clips and verify the shelves are level.
CVO-10
1.
Install the first shelf in the bottom-most pair of shelf slides, just above the temperature sensor
probe. A shelf must be installed here for the oven to accurately control the chamber temperature.
2.
Slide the remaining shelves into the shelf slides.
Temperature Sensor Probe
1
st
Shelf
Side Cutaway View
Shelf Clip
Rocking Motion
Probe
Shelf
Summary of Contents for CVO-10
Page 1: ...Vacuum Purge Ovens Installation Operation Manual CVO 5 CVO 5 2 CVO 10 ...
Page 4: ...4 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Page 10: ...10 P a g e cascadesciences com Tel 503 847 9047 UNIT SPECIFICATIONS ...
Page 18: ...18 P a g e cascadesciences com Tel 503 847 9047 RECEIVING YOUR OVEN ...
Page 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Page 52: ...52 P a g e cascadesciences com Tel 503 847 9047 USER MAINTENANCE ...
Page 55: ...55 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS ...
Page 56: ......