When selecting a location to install the oven, consider all environmental conditions that can affect
effective temperature range, uniformity, and stability of the oven. For example:
•
Other ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling ducts, or other sources of fast moving air currents
•
High-traffic areas
•
Direct sunlight
This oven is intended for use indoors, at room temperatures between
15
°
C and 40
°
C (59
°
F and 104
°
F)
,
at no greater than
80% Relative Humidity
(at 25
°
C / 77
°
F). Operating the unit outside of these conditions
may adversely affect its temperature range and stability.
R
EQUIRED
C
LEARANCES
These clearances are required to provide sufficient air flows for ventilation and cooling.
•
Allow at least
12 inches (300 mm)
of empty space between the top and back of the oven and
any walls or partitions for unobstructed airflow and cooling.
•
Allow at least
6 inches (150 mm)
of ventilation on the sides.
•
Do not place objects on top of the oven
.
•
Allow at least
6 inches (152 mm)
from the fan vent on the back of the oven to the nearest
wall or partition. Keep the fan unobstructed at all times.
•
A KF-25 vacuum port is located on the back of the oven for introducing probes into the
chamber or connecting to an external vacuum supply source. Leave sufficient clearance for
end-users to safely access this port if you will be using it.
6” (150 mm) Fan
KF-25 Port
6” (150 mm)
Door Swing
12” (300 mm)
12” (300 mm)
6” (150 mm)
12” (300 mm)
20” (521 mm)
Summary of Contents for CVAC-9.3
Page 1: ...Installation Operation Manual CVAC 9 3 Vacuum Oven ...
Page 5: ...5 P a g e cascadesciences com Tel 503 847 9047 REPLACEMENT PARTS 53 ...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS ...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS ...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 INTRODUCTION ...
Page 28: ...28 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION ...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 GRAPHIC SYMBOLS ...
Page 34: ...34 P a g e cascadesciences com Tel 503 847 9047 CONTROL OVERVIEW ...
Page 46: ...46 P a g e cascadesciences com Tel 503 847 9047 OPERATION ...
Page 54: ......