Summary of Contents for CI-2400BS

Page 1: ......

Page 2: ...CI 2400BS Weighing Indicator OWNER S MANUAL ...

Page 3: ...PPEARANCE 7 FRONTPANELDESCRIPTION 8 REARPANELDESCRIPTION 10 SEALINGMETHOD 14 SYSTEMMODE 15 GENERALFUNCTION DESCRIPTION 23 SETMODE 26 TESTMODE 34 CALIBRATIONMODE 37 SIMPLECALIBRATIONMODE 40 ONLYZEROCALIBRATION 42 GRAVITYCOMPENSATION THEOTHERSSETMODE 44 OPTIONS 48 ERRORMESSAGE TROUBLESHOOTING 50 ...

Page 4: ... the indicator on a flat and stable surface Do not use inflammable materials in cleaning Do not severely press because the light pressing of keys can incite the operation Keep the main body from rain and keep in dry area ...

Page 5: ...4 Do not subject the scale to sudden temperature changes Keep the scale away from strong EMI noises may cause incorrect weight readings ...

Page 6: ...may use this device to its full potential 1 Features Water proof IP65 Appropriate for weight and measurement system Easy operation and various functions Simple calibration Full digital calibration installing the automatic weight at a time Watchdog circuitry System restoration The construction of stainless steel Standard fixed bracket type 2 Main Function Adjustable display rate Digital filter func...

Page 7: ...nter Option 1 Option 2 Option 3 Option 4 Option 5 Option 6 3 Option TECHNICAL SPECIFICATION DC 5V up to 6 load cells 100mA 0 05mV 5 m V 2 V D NTEP OIML 0 5 V D Non NTEP OIML 0 01 F S 1 200 000 1 10 000 Max 10 times sec Load cell excitation voltage Zero adjust range Input sensitivity Nonlinearity A D internal resolution Analog Part A D Conversion ...

Page 8: ...table Hold AC 110V 220V DC 12V 850mA 1 2 W 10 C 40 C 14 F 104 F 2 0 0 m m 1 3 0 m m 5 3 m m Approx 1kg 2 2lb Span calibration Display Size of letter Display below zero Status display Power P o w e r consumption Temperature Digital Part MEASURE OF APPEARANCE Fixed Bracket Type CI 2400BS Standard 53 70 200 240 130 165 CI 2400BS CAPACITY ...

Page 9: ...any data is printed HOLD lamp ON when you weigh the moving object BAT lamp In case of using rechargeable battery it is displayed the battery recharging time POWER lamp ON when it is connected to power lamp ON when the current weight exceeds normal weight lamp ON with alert when the current weight is normal lamp ON when the current weight is 90 100 of normal weight lamp ON when the current weight i...

Page 10: ...s weight is NET weight Used to change weight units kg lb oz to weigh the moving object Auto Manual Used to power ON OFF Key functions in COUNT mode Used to display the unit weight for approx 5 sec Used to display the current weight for approx 5 sec Key functions in SET mode Digit input method Input the digit with using the below keys key Change the set value Increase the first place value to 1 key...

Page 11: ...est To Send 5 Clear To Send 6 Data Set Ready 8 Carrier Detect 20 Data Terminal Ready Connect RS 232C Port of Computer 25pin port Female RS 232C port of Computer RS 232C port of CI 2400BS RXD TXD GND 3 Transmit Data 2 Receive Data 5 Chassis Ground 1 Carrier Detect 4 Data Terminal Ready 6 Data Set Ready 7 Request to Send 8 Clear to Send 9pin port Female RS 232C port of Computer RS 232C port of CI 24...

Page 12: ... allow the receiver side to selectively receive information supplied from the indicator Example When host transmit device ID 01 ASCII code the indicator of device 01 answers Device ID is able to set at F10 Data 8 bytes Weight data with decimal point 1 13 5 kg 0 0 0 0 1 3 5 2 135 kg 0 0 0 0 1 3 5 0 3 135 kg 0 0 0 1 3 5 0 Simple Transmit Receive Program Language BASIC 10 OPEN COM1 9600 N 8 1 As 1 20...

Page 13: ...tatus DONE FALSE bioscom 0 SETTINGS COM1 cprintf BIOSCOM ESC to exit n while DONE status bioscom 3 0 COM1 if status DATA_READY if out bioscom 2 0 COM1 0 7F 0 putch out if kbhit if in getch 1B DONE TRUE bioscom 1 in COM1 return 0 Connection FS 7000D Printer RXD TXD GND 2 Transmit Data 3 Receive Data 14 Signal Ground 15pin port Female Serial port of FS 7000D Printer RS 232C port of CI 2400BS ...

Page 14: ...nt In that case please note the following diagram Manufacturer s wire colors N0 1 No 2 No 3 No 4 No 5 EX EX SIG SIG GND Red White Green Blue Shield Red Black Green White Shield Red Black Green White Shield Red Black Green White Shield Green Black White Red Yellow Red Blue White Black Shield Red Black Green White Shield Red White Green Blue Yellow Red Black White Green Shield Red Blue White Black Y...

Page 15: ...resistor SEALING METHOD Perform the sealing as shown in the below figure after the calibration is completed Screw CAL switch sealing bolt Connect to the sealing wire as the follows and compress the sealing lead Cal Switch Sealing Bolt Sealing Bolt Sealing Lead How to Enter CAL Mode TEST Mode Turn on the power while pressing the CAL switch on the back side How to Enter SET Mode Turn on the power wh...

Page 16: ...nd it is displayed the following message Execute depending on the below mode conversion and setting method 2 Digit Input Method Input the digit in use of the below keys key Used to escape from mode key Used to increase the first place of value by one key Used to set current value 10 key After completing the input it is moved to next mode key Used to set the inputted value to 0 Conversion method As...

Page 17: ...6 Press one Set input value to 0 Press twice Input 2 as the first place value Press once Used to set current value 10 Press once Used to set current value 10 Press five times Input 5 as the first place value ZERO MODE Note The decimal point automatically is set to location which is set in CAL mode ...

Page 18: ...tion Step 1 Step 2 Step 3 Empty Press once Press once It is operated generally 4 High Low Comparing Mode Conversion In case of no resetting the High Low value Display Key Platform Description Step 1 Step 2 Step 3 Empty Press twice Press once It is operated generally ...

Page 19: ... Low value Display Key Platform Description Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Empty Press twice Press once Input low weight value Refer to digit input method Input high weight value Refer to digit input method ZERO MODE ...

Page 20: ...2 Step 3 Empty Press there times Press once It is operated generally Display Key Platform Description Step 1 Step 2 Step 3 Step 4 Step 5 Press three times Press once Display the set weight Input unit weight MODE MODE After completi ng the digit input In case of setting the unit weight Min range of unit weight Min Div ...

Page 21: ...mum division is 0 1kg in CAL mode you can input to 0 00kg which increased a digit place than minimum division In case of setting the unit weight by sample Samples 10 200 10 unit Display Key Platform Description Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Press three times Press once Press once Press once Press five times Input unit 10pcs Press once with loading Under setting the MODE MODE Emp...

Page 22: ... Description Step 1 Step 2 Step 3 Press four times Press once It is operated generally Note 1 In case of setting the unit weight by sample count the min unit weight shall be more than min division value Note 2 It can be input to 200pcs as 10pcs of input unit 6 Count Comparing Mode Conversion In case of no setting the High Low quantity ...

Page 23: ...ity Display Key Platform Description Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Press four times Press once Low quantity setting Input low quantity Press once High quantity setting Input high quantity MODE Press key Press key MODE Empty ...

Page 24: ... capacity Note The weight including TARE weight can t exceed the maximum capacity In case to want to know the GROSS weight In case to want to know the NET weight Zero point changed Press ZERO key Zero compensation Put the TARE on platform Press TARE key The TARE weight is Press NET GROSS key Display object weight Remove the TARE and object from platform And it will be displayed Press NET GROSS key...

Page 25: ...lease It is executed at emptying the platform Empty the platform Press HOLD key AH ON message is displayed Empty the platform Press HOLD key After showing AHOFF message for 1 sec it is normally The Hold lamp is ON Put the object 10kg on Press HOLD key Hold message is displayed for The weight is averaged for When the weight change i s stable after putting the object 20kg on The weight is averaged T...

Page 26: ...current net weight Print format is decided by F15 of set mode The weighing number from 1 to 200 2 SUM Print It can be printed normally in selecting 1 or 2 of F13 If you press LIGHT PRINT key for 2 sec it will be printed as the follows After performing Sum print function the weighing number and sum value is automatically released After power turning on the weighing number and sum value is automatic...

Page 27: ... input value LIGHT PRINT Used to escape from SET mode 3 Set Menu GENERAL FUNCTION F1 Set weight unit F2 Auto power off F3 LIGHT PRINT key usage F4 Weight backup F5 Auto zero compensation range F6 Digital filter RS232C COMMUNICATION FUNCTION F10 Designation of serial interface baud rate F11 Designation of serial interface method F12 Device ID PRINTER FUNCTION F13 Using printer set F14 Auto manual p...

Page 28: ...wer OFF is not used F2 01 Power off when the weight is not changed or no key pressing for 3 min F2 02 Power off when the weight is not changed or no key pressing for 5 min F2 03 Power off when the weight is not changed or no key pressing for 10 min F02 SET 00 03 GENERAL FUNCTION Note Use this function depending on unit of weights which is used in CAL mode Note This function performs when the weigh...

Page 29: ...ht backup DISPLAY DESCRIPTION F4 00 Weight backup OFF F4 01 Weight backup ON F04 SET 00 01 FUNCTION Automatic zero compensation DISPLAY DESCRIPTION F5 00 No compensation F5 05 Compensation if it is gradually changed below 5 digit for 2 sec F5 09 Compensation if it is gradually changed below 9 digit for 2 sec F05 SET 00 09 Note 1 In case occurring sudden power failure it can be memoried the moment ...

Page 30: ...IPTION F10 00 2400 F10 01 4800 F10 02 9600 F10 03 19200 F10 SET 00 03 RS232C COMMUNICATION FUNCTION Output method RS232C data DISPLAY DESCRIPTION F11 00 No data output F11 01 Output data in state of stable unstable F11 02 Output data only in stable state F11 03 Output data when data is required F11 SET 00 03 Note 1 The default value is set to 0 at factory Note 2 It is available in case of setting ...

Page 31: ...utomatic print when the weight is stable or you press the key it will be printed F14 SET 00 01 FUNCTION Print form DISPLAY DESCRIPTION F15 00 Form 0 F15 01 Form 1 F15 02 Form 2 F15 SET 00 02 Form 0 Date Time Weighing No Net weight 1999 1 1 13 15 No 001 50 0kg No 002 100 0kg No 003 200 5kg Form 1 Date Time Net weight 1999 1 1 13 15 NET 50 0kg NET 100 0kg NET 200 5kg Form 2 Date Time Weighing No Net...

Page 32: ...c F16 SET 00 15 AUTOMATIC HOLD FUNCTION FUNCTION Set initializing division Only available in automatic hold DISPLAY DESCRIPTION F17 01 If it is swang over 2 division it will be initialized F17 07 If it is swang over 14 division it will be initialized F17 15 If it is swang over 30 division it will be initialized F17 SET 01 15 FUNCTION Relay mode DISPLAY DESCRIPTION F20 00 Limit mode F20 01 Limit ty...

Page 33: ...OUT 1 If weight is under 5 division this will be turned on Note 2 LOW HIGH RELAY OUT 2 3 4 When weight is over 5 division comparing with setting weight value and displaying weight value it will be turned ON OFF ZERO OUT 1 LOW OUT 2 NORMAL OUT 3 HIGH OUT 4 WEIGHT RELAY 0kg 80kg LOW limit HIGH limit 100kg F2001 Limit Type Checker Mode ON ON ON ON OFF OFF OFF OFF ...

Page 34: ...1 1st day C3 SET 01 31 FUNCTION Set time DISPLAY DESCRIPTION C4 00 AM 00 hour C4 SET 00 23 FUNCTION Set minute DISPLAY DESCRIPTION C5 59 59 minute C5 SET 00 59 FUNCTION Set second DISPLAY DESCRIPTION C6 59 59 second C6 SET 00 59 Note 1 If you press Enter key after completing set mode it will be moved to the initial menu of Set Mode Note 2 Press LIGHT PRINT key it will be moved to the Normal Mode ...

Page 35: ...t menu 3 TEST Menu TEST 1 LCD Screen Test TEST 2 A D Converter Test TEST 3 Key Test TEST 4 Serial Communication Test TEST 1 Function LCD Screen Test Key Display Description Key Move to the next menu The mode and key are displayed Press ZERO key and then it will be moved to Test mode TEST 1 condition All lamps on LCD SCREEN is turned on for testing Press ZERO tESt Tare Cal ...

Page 36: ... menu Others Key Test execution TEST 3 condition Press the key to be tested and the No of key mode should be identify with code of key as the follows Note Check whether digital value is changing If the digital value is fixed or zero is displayed please check the connection of load cell TEST 3 Function Key Test Note If you press UNIT HOLD key after checking key number it will be automatically moved...

Page 37: ...e Transmitted 1 Received 2 Transmitted 1 Press When this test is completed it is automatically moved to NORMAL MODE Note 1 Do this test after the connection of serial port of computer and serial port of indicator is done Note 2 Send No 2 in computer keyboard and check if indicator receives No 2 Send No 1 in indicator keyboard and check if computer receives No 1 Note 3 Do this test after baud rate ...

Page 38: ...AL 1 CAL 5 CAL 1 Maximum Capacity CAL 2 Minimum Division CAL 3 Setting Weight CAL 4 Zero Calibration CAL 5 Span Calibration CAL 1 Function Maximum Capacity Set Range 1 99 999 Key Display Description Key Control division Key Control digit Key Store and move to the next menu CAL 1 condition Existing maximum capacity Setting maximum capacity Press key ex Note 1 The maximum capacity means the maximum ...

Page 39: ...on Note 2 External resolution is obtained by dividing the min division by the maximum capacity Set the resolution to be within 1 10 000 Note 3 The maximum capacity minimum division is over 10 000 Error message CH 11 will occur Note 4 If you press key it will be moved to CAL 3 CAL 3 Function Setting Weight in Span Calibration Range 1 Maximum Capacity of CAL 1 Maximum capacity value of CAL 1 Setting...

Page 40: ...ed from CAL mode Press key CAL mode is completed Move to the initial menu automatically Note 1 After completing zero calibration it will be automatically moved to CAL 5 Note 2 If the zero value is too low or high CH 13 will occur You lower the resolution and reset the zero calibration CAL 5 Function Span Calibration Note 1 END message is displayed and move to the initial menu when span calibration...

Page 41: ... be moved to simple calibration mode And then it will be started from zero calibration Function Zero Calibration Key Display Description Key Move to the next menu Empty the platform and when LCD screen shows ULOAD Press key Under zero calibration After completing zero calibration it will be automatically moved to span calibration Note If the zero value is too low or high CH 13 will occur You lover...

Page 42: ...de Press key CAL mode is completed Move to the initial menu automatically Function Span Calibration ZERO tESt Tare Cal Note 1 END message is displayed and move to the initial menu when span calibration is completed Note 2 If the span value is too low or high CH 13 will occur You lower the resolution and reset the span calibration Note 3 After completing CAL mode it will be moved to initial menu ...

Page 43: ...e back side Press TARE key and UNIT HOLD key it will be moved to calibration mode Display Key Platform Description Step 1 Step 2 Step 3 Step 4 Step 5 Displaying maximum capacity Displaying minimum division Displaying the setting weight Press once ZERO MODE Empty Empty Empty Empty ...

Page 44: ...libration i s completed It is escaped from zero calibration CAL mode is completed Move to the Empty Empty Empty zero tESt Tare Cal Note 1 If the zero value is too low or high in STEP 6 CH 13 will occur Note 2 If you find CH 14 check the platform and the connection of load cell and do this test again ...

Page 45: ...ws it to be calibrated in one location and then adjusted to match the acceleration of gravity at another location where it will be used Multi interval Dual range EX 1 If you calibrated minimum division to 1kg and maximum capacity to 3000kg 0 1500kg minimum division is 0 5kg 1500 3000kg minimum division is 1kg EX 2 If you calibrated minimum division to 2kg and maximum capacity to 5000kg 0 2500kg mi...

Page 46: ...GRCAL condition 9798 9 798m s2 9790 9 790m s 2 Press key SET 9700 9900 Unit m s2 GRCAL FUNCTION Gravity acceleration of using location DISPLAY DESCRIPTION GRSET GRSET condition 9798 9 798m s2 9778 9 778m s 2 Press key SET 9700 9900 Unit m s2 GRSET Note If you press key it will be moved to GRSET mode Note 1 If the GRCAL value is same with GRSET it doesn t need to do gravity compensation set Note 2 ...

Page 47: ...wait 9 793m s 2 Lisbon 9 801m s 2 London Greenwhich 9 812m s 2 Los Angeles 9 796m s 2 Madrid 9 800m s 2 Manila 9 784m s 2 Melbourne 9 800m s2 Mexico City 9 779m s 2 Milan 9 806m s 2 New York 9 802m s 2 Oslo 9 819m s 2 Ottawa 9 806m s 2 Paris 9 809m s 2 Rio de Janeiro 9 788m s 2 Rome 9 803m s 2 San Francisco 9 800m s 2 Singapore 9 781m s 2 Stockholm 9 818m s 2 Sydney 9 797m s 2 Taichung 9 789m s 2 ...

Page 48: ...al point to S3 1 Use decimal point to Press key Gravity compensation the others set mode are completed When this test is completed it is automatically moved to NORMAL MODE SET 0 1 COMMA Note 1 In case of setting S1 0 standard division set in CAL mode is used maximum capacity Note 2 In case of setting S1 1 the following division is used in the following range 0 1 2 of MAX capacity one lower divisio...

Page 49: ...necting method of RS 485 422 PORT OPTION 1 RS 422 485 Serial Interface IN IN OUT OUT Transmit Data Transmit Data Receive Data Receive Data Serial port of Computer RS485 Port of CI 2400BS IN IN OUT OUT 6 Transmit Data 4 Transmit Data 5 Receive Data 7 Receive Data Serial port of Sub display RS485 Port of CI 2400BS ...

Page 50: ...000mAh battery 2 Ni Cd Battery DC 7 2V 2000mAh battery pack 3 Install Connect output port of Ni Cd battery pack to CN3 of charge board Connect output port of adapter to CN1 of charge board after confirming polarity of adapter Connect CN2 of charge board to CN1 of CI 2400BS main board 4 Method If adapter code is connected to AC power to charge the battery power lamp is ON When charging battery char...

Page 51: ...lace without shaking and power on CH 04 Reason The initial zero range exceeds above 10 of maximum capacity Trouble shooting Check the platform CH 05 Reason Pressing any key for long time or having problems in key part Trouble shooting If it can t be found problems in key part please call A S noprt Reason When it is setting F13 0 or when you press print key to print current and unit weight at count...

Page 52: ... capacity Trouble shooting Set span setting weight to 10 100 range of maximum capacity in CAL 3 of CAL mode CH 13 Reason Load cell output voltage span value is too high at span calibration Trouble shooting Check if weighing platform is in contact with something Or set external resolution to lower external resolution CH 14 Reason Load cell output voltage zero value is too high or too low at zero ca...