Refer to the Dreal™ system Instructions for Use (IFU) for

detailed instructions on motor and irrigation setup

d

rea

d

rea

OR Setup Guide

50cc

or larger

IV bag

Motor

handpiece

(foot operated)

Refer to Model Selection
Guide for advice on the
proper tip configuration.

Must be sterilized
prior to use.

Standard

Sterile tubing

Luer

lock

Roller-

clamp

Should be placed in the 
sterile field, within the 
surgeon’s reach; Dripping 
rate of ~1drop/sec 
provides enough 
irrigation to keep the tip 
from heating and allows 
debris to wash away.

Length should allow the 
surgeon to work 
conveniently or an 
extension may be used.

The Dreal™ is intended to cut bone in neurosurgical and spinal applications. 
Prior to use, the surgeon should refer to the product Instructions for Use (IFU)

 

for complete indications, warnings, precautions and contraindications. 

Official US Distributor

Summary of Contents for Dreal 2L75FU

Page 1: ...k Roller clamp Should be placed in the sterile field within the surgeon s reach Dripping rate of 1drop sec provides enough irrigation to keep the tip from heating and allows debris to wash away Length...

Page 2: ...stems cleared for Neuro and Spine surgery in the US include Stryker CORE TPS with Micro Drill or Rem B handpieces Medtronic Midas REX Legend with Carevature Adaptor DePuySynthesAnspach Xmax MicroMaxPl...

Page 3: ...1 Instructions For Use For DrealTM...

Page 4: ...g or re sterilization of the device is strictly prohibited as reuse of the device can compromise the DrealTM performance characteristics DO NOT use the DrealTM without active irrigation to prevent tis...

Page 5: ...however it can rotate only clockwise right hand or Forward NOTE the Dreal was tested for compatibility with the following motor systems Stryker Consolidated Operating Room Equipment CORE Motor Disinf...

Page 6: ...power On Off plug on the control unit should NOT be touched by the gloved hand of the health care provider NOTE make sure that the motor is set to clockwise right hand or Forward direction Set the re...

Page 7: ...tep by step approach Please refer to the Indications for Use section at the beginning of this document In summary the Dreal is intended to cut bone in procedures such as laminectomy laminotomy and cra...

Page 8: ...5 Cavity is drilled without nerve root manipulation In case where there is no ligamentum e g in the cranium or if it has been removed retraction of the neural structures may be done if necessary Tip...

Page 9: ...r local authorized representative Packaging The Dreal package is sterile as indicated by its label and you should verify the package integrity when receiving the Dreal and before you open each Dreal p...

Page 10: ...ly Do not re sterilize Do not use if package is damaged or opened Sterilized by Gamma Consult Instruction For Use enclosed with the DrealTM Temperature limitation upper and lower limits of temperature...

Page 11: ...3S45FUS 3S45FUD 3S45FI 3L75FU 2L75FU 3L10DS 3S45FU 3S45FUS 3S45FUD 3S45FI 3L75FU 2L75FU 3L10DS 3mm 3mm 3mm 3mm 3mm 2mm 3mm Diameter 45 45 45 45 75 75 10 Tip angle 10mm 10mm 10mm 10mm 18mm 18mm 6mm Mi...

Page 12: ...pproach 3S45FU 3S45FUS 3S45FUD 3S45FI 3L75FU 2L75FU 3L10DS Posterior Osteophyte removal lumbar Anterior Cervical foraminotomy uncusectomy Posterior Lateral Decompression Below the root Posterior Later...

Page 13: ......

Reviews: