![Carestream DRYVIEW 5950 User Manual Download Page 41](http://html1.mh-extra.com/html/carestream/dryview-5950/dryview-5950_user-manual_2557844041.webp)
Maintenance and Troubleshooting
AA2204_en | 2012-07-27
3–13
Job Manager
Related topics
Load a Different Film Size to Match a Print Request
Job
Manager
Status
Display Screen
Web Portal
Description
Active
Not Applicable
Shows how
many print
requests have
been
initiated.
The imager is accepting DICOM job
requests and film is available for all
current jobs.
No Media
Power symbol is
green
Film Size symbol is
yellow
The required film size
is shown on the
display screen
Shows how
many jobs
with this
status are in
the job
queue.
The imager is accepting DICOM job
requests, but film of the correct size
and type is not available for at least
one current job.
Offline
Status code: 704
Power symbol is
green
DICOM
Offline
The imager cannot accept any DICOM
job requests. Restart the imager.
SHORT TERM CONFIDENTIAL