IVAC® PCAM® Syringe Pump

Directions For Use
en

Summary of Contents for IVAC PCAM

Page 1: ...s IVAC PCAM Syringe Pump Directions For Use en...

Page 2: ...Symbol Definitions 5 Operating Precautions 7 Getting Started 9 Basic Features 12 Alarms 17 Configured Options 18 Configured Options Record 21 Drugs and Protocols Record 22 Specifications 23 Maintenan...

Page 3: ...epidural and subcutaneous routes Supporting fluid therapy blood transfusions and parenteral feeding The pump is compatible with a wide range of standard single use disposable Luer lock syringes It acc...

Page 4: ...1000DF00451 Issue 6 3 34 Features of the Pump PLUS MINUS RS232 Connector Display Handset Connector LOCK 1 START STOP PURGE ARROW HISTORY PRINT COVER LOCK 2 on side of case...

Page 5: ...ROW buttons Use as softkeys in conjunction with the prompts shown on the display OFF SET RUN LOCK 1 Insert key into LOCK 1 keyswitch and turn key to switch between OFF SET and RUN positions OFF Turns...

Page 6: ...ment k Type CF applied part Degree of protection against electrical shock P Protected against splashing fluid degree of protection against fluid ingress r Alternating Current s Device complies with th...

Page 7: ...icon When enabled this icon is shown on the Display It provides a visual indicator of current pumping pressure and pressure level at which the alarm will operate 4h MAXIMUM DOSE icon When enabled this...

Page 8: ...extension set and patient connections and follow the priming procedure specified herein Operating Environment When using any infusion pump in conjunction with other pumps or devices requiring vascula...

Page 9: ...unity This pump is a CISPR 11 Group 1 Class A device and uses RF energy only for its internal function in the normal product offering Therefore its RF emissions are very low and are not likely to caus...

Page 10: ...he pump in a choice of four positions A Do not mount the pump with the AC power inlet or the syringe pointing upwards This could affect the electrical safety in the event of a fluid spill or lead to t...

Page 11: ...ze the finger grips on the plunger holder and slide the mechanism to the right until the syringe finger flanges locates in the V slot A Gently advance the syringe until the finger flanges touch the fr...

Page 12: ...ress OK 7 CONFIRM SYRINGE Check that the syringe type and size being used matches display If required the make of syringe can be changed by pressing the CHANGE TYPE softkey Press OK 8 PURGE if require...

Page 13: ...ion for example during the PCA lock out period The clinician over ride is a special feature which can be configured according to the specific clinical situation Clinician over ride can also be used in...

Page 14: ...helps to indicate if the PCA protocol needs modification and when to end treatment Record of the total amount of drug administered to the patient over the last 24 hours The latest hour is shown at th...

Page 15: ...header Event Log Press J button until event log is displayed Use S buttons to position display at start point for events to be printed Press V button All events will be printed from information on sc...

Page 16: ...ING DRUG NAME WILL ALTER SAFETY LIMITS FOR PROTOCOL CONTINUE CONFIRM DRUG Y To change to another Pre Set Protocol Press NEXT PROTOCOL to display the step through the pre set protocol Pre set protocols...

Page 17: ...s 1 43ml 715 g with the remaining 0 0099ml being carried over into the display of the next demand The next demand would be displayed accumulatively as 2 87ml 1 4399ml 1 4399ml 2 8798ml with the remain...

Page 18: ...nd after 30 minutes a continuous audible alarm will indicate that the battery is exhausted Reconnect to AC power supply to continue operation and charge internal battery BATTERY EXHAUSTED BATTERY EXHA...

Page 19: ...ll back will be cancelled for up to 2 mins or extended to 15 mins To extend call back alarms press and hold the X button for 4 seconds 6 Display Sleep YES display goes blank after 2 minutes NO display...

Page 20: ...Mass Vol Modes YES Allow a mix of mass and volume dose mode for drugs and protocols NO All drugs and protocols use mass dose mode A To minimise the risk of incorrect confirmation of the syringe type i...

Page 21: ...eneral Options 7 MIN DRUG CONC Use R Q buttons to set minimum concentration For volume mode concentration can be set to OFF the lowest value Press OK when complete 8 MAX DRUG CONC Use R Q buttons to s...

Page 22: ...INFUSIONS YES NO YES LOADING DOSES YES NO YES MAX DOSE LIMITS YES NO YES VARIABLE DOSE RATES YES NO YES COMMS PUMP IDENTITY 000 127 001 COMMS ENABLED YES NO YES NURSE CALL YES NO NO NURSE CALL INVERT...

Page 23: ...0 180 minutes 0 180 minutes Minimum PCA Dose 0 g 99 9mg 0 0ml 99 9ml Maximum PCA Dose 0 g 99 9mg 0 0ml 99 9ml Maximum Continuous 0 g h 999 0mg h 0 0ml h 35ml h Maximum Loading Dose 0 g 99 9mg 0 0ml 9...

Page 24: ...Specifications Operating Temperature 10 C 40 C Operating Relative Humidity 30 75 Operating Atmospheric Pressure 700hPa 1060hPa Transport Storage Temperature 20 C 50 C Transport Storage Relative Humid...

Page 25: ...the worst case values for line pressure time to alarm and bolus volume that can be expected in the event of an occlusion when the IVAC 50ml syringe and G40020B extension set are selected Alarm Level...

Page 26: ...Use supplied with the extension set prior to use Standard Sets G40015 Standard PVC Syringe Extension Set 150 cm Priming Volume 2 6ml G40020B Standard PVC Syringe Extension Set 200 cm Priming Volume 1...

Page 27: ...with clamp 152 cm Priming Volume 0 5ml 30832 PVC Y Syringe Extension Set with back check valve and 2 clamps 178 cm Priming Volume 1 5ml 30842E PVC Syringe Extension Set with back check valve SmartSit...

Page 28: ...ill not be responsible should any of these actions be performed outside the instructions or information supplied by CareFusion Refer to the Technical Service Manual for the access code for technical s...

Page 29: ...n a clean dry atmosphere at room temperature and if available employ the original packaging for protection Once every 3 months during storage carry out functional tests as described in the Technical S...

Page 30: ...ies with the user of the equipment This software should include detection of the disconnection or other failure of the RS232 cable The protocol is detailed in the Technical Service Manual and is for g...

Page 31: ...t of the curve As the observation window is reduced short term fluctuations have greater effects as represented by the mouth of the trumpet Knowledge of system accuracy over various observation window...

Page 32: ...oject Once 2 minute minimum maximum 5 5 0ml h Long term mean accuracy 2 is in addition to the trumpet curve percentage error Performance with syringes IVAC Zeneca Terumo Nipro Fresenius B Braun Perfus...

Page 33: ...ederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 902 555 661 Fax 31 0 30 2...

Page 34: ...at www carefusion co uk alaris technical A username and password are required to access our manuals Please contact local customer services representative to obtain login details Part Number Descripti...

Page 35: ...1000DF00451 Issue 6 34 34 Page Intentionally Left Blank...

Page 36: ...n of CareFusion Corporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction dis...