Alarms
•
High airway pressure – 3 cmH20 above measured airway pressure
•
High circuit pressure – maximum 11 cmH20 in time triggered Biphasic mode
•
High circuit pressure – maximum 15 cmH20 in patient triggered Biphasic tr
mode (Comprehensive only. The comprehensive configuration is not
available for sale in the United States.)
•
Low airway pressure – 2 cmH20 below measured airway pressure or 1
cmH20 if otherwise would be zero
•
High and Low delivered Oxygen concent5% of setting
•
Low breath rate alarm
•
Low or Flat battery charge level
•
Alarm volume (electronic alarms) 70 dBa at 1 meter
Pneumatic Supply
•
Patient Gas Outlet:
15 mm standard taper fitting
•
Patient Pressure Input:
4.5 mm Luer taper fitting
•
Gas Supply:
Nominal 4 bar or 60 psi, clean, dry medical air and oxygen
•
Range:
2.8 to 6 bar (40 to 90 psi); maximum differential pressure 2 bar
(30 psi)
•
Manometer:
Range 0 to + 20 cmH
2
O, accuracy,
±
2% of span
•
Gas Connections:
Standard DISS or NIST connectors
Electrical Supply
•
Input Voltage:
100-230 VAC
•
Input Frequency:
50/60 Hz
•
Power Consumption:
50 VA maximum
•
Fuse Rating For 220 V nominal operation:
“T” Type 2.5 A at 250 V
•
Device Housing Protection rating level:
IPX1
•
Battery Working Time:
2 hours (from fully charged state)
•
Battery Charging Time:
max. 16 hours
Summary of Contents for Infant Flow LP SiPAP
Page 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Page 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Page 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Page 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Page 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...