26
Servicing
If your unit requires service or repair please see page 37 for contact
details. A full service manual including circuit diagram and parts list
is available upon request.
Trouble Shooting Information
Should you encounter problems operating your BabyCO meter,
please consult the table below:
Problem
Possible Cause
Solution
BabyCO cannot be
switched on or “bat”
is displayed
Batteries are flat
Change the battery
Battery life is
shorter than
expected
Unit not being
switched off
Turn the unit off after
use
“CEL” is displayed
Fuel cell is
depleted
Return unit for cell
replacement
“Err” is displayed
Gas cylinder empty Check valve on cylinder
for contents and replace
cylinder if necessary
Check the calibration
value on the cylinder is
20ppm.
“bt2” is displayed
Internal battery
has expired
Return unit for battery
replacement
“gAS” is displayed
Fuel cell is
depleted
Return unit for cell
replacement
Fuel cell is
contaminated by
residual gas or
solvents
Allow unit to be
exposed to clean air.
Turn unit ON and await
message to disappear,
if not return unit for cell
replacement.
Summary of Contents for BabyCO Meter
Page 2: ......
Page 10: ...8 ...
Page 41: ...39 ...
Page 42: ...40 ...
Page 47: ...45 6 5 4 1 2 3 CO FCOHb OFF ...
Page 50: ...48 Das BabyCO ist zur Verwendung durch Ärzte und medizinisches Fachpersonal vorgesehen ...
Page 79: ...77 ...
Page 80: ...78 ...
Page 114: ...112 ...
Page 146: ...144 ...
Page 181: ...179 ...
Page 182: ...180 ...
Page 216: ...214 ...
Page 221: ...219 6 5 4 1 2 3 CO FCOHb OFF ...
Page 254: ...252 ...
Page 264: ...262 将滑动开关推至 CO PPM 位置 等待设备显示清零 如下所示 将控制阀紧紧拧到气缸上 并连接气源 串行 定标 开关 ...