18
breath holding countdown timer to be configured to individual
requirements.
Note
: The BabyCO should only be connected to a computer that is
manufactured in accordance with EN 60601-1.
Note:
Keep the PC out of reach of the patient at all times.
Calibration
Calibration will remain stable to within 2% over one month and
typically to within 10% over 6 months. CareFusion supplies
calibration gas (20ppm CO in air) and recommends that the unit be
recalibrated on a 6 monthly basis. See page 35 for calibration
accessories.
The gas is supplied in convenient, disposable, aluminium aerosol
cans containing 20 litres of gas (Cat. No.36-MCG020).
To carry out the calibration locate the calibration switch on the right-
hand side of the instrument as shown below.
Serial
port
Calibration
Switch