17
moisture has evaporated from the surface of the sensor
before reuse.
If the unit is switched on again too quickly after use there may be a
response to residual expired carbon monoxide from the previous
test.
In this case the display will show:
If this is displayed then turn the unit off, remove the mouthpiece
adapter, and expose to ambient air for 2 minutes before repeating
the test.
Note:
If this warning appears again after following the above
procedure then turn off and leave the sensor exposed to ambient air
for a further 3 minutes. If the same message appears again then
this indicates possible contamination of the sensor with a solvent.
In this case remove all sources of solvent from the vicinity of the
sensor and expose to ambient air for 24 hours before switching on
again.
PC Connection
The BabyCO meter may be connected to the serial port of a PC
running COBRA (CO Breath Analysis) software.
This software allows the measurements to be read by a PC and
automatically entered onto a pre-defined report for subsequent
printing and filing. It also allows the CO level indicators and the
Summary of Contents for BabyCO Meter
Page 2: ......
Page 10: ...8 ...
Page 41: ...39 ...
Page 42: ...40 ...
Page 47: ...45 6 5 4 1 2 3 CO FCOHb OFF ...
Page 50: ...48 Das BabyCO ist zur Verwendung durch Ärzte und medizinisches Fachpersonal vorgesehen ...
Page 79: ...77 ...
Page 80: ...78 ...
Page 114: ...112 ...
Page 146: ...144 ...
Page 181: ...179 ...
Page 182: ...180 ...
Page 216: ...214 ...
Page 221: ...219 6 5 4 1 2 3 CO FCOHb OFF ...
Page 254: ...252 ...
Page 264: ...262 将滑动开关推至 CO PPM 位置 等待设备显示清零 如下所示 将控制阀紧紧拧到气缸上 并连接气源 串行 定标 开关 ...