Alaris™ VP Plus Guardrails™ Volumetric Pump
Flow Sensor Operation (Optional)
1000DF00660 Issue 9
34/48
Flow Sensor Operation (Optional)
A
The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause
the Pump to alarm if a significant deviation from the infusion rate occurs. The flow sensor will also be able to detect
empty containers. For this reason we recommend use of a flow sensor wherever possible excluding secondary
infusions.
IVAC® Model 180 Flow Sensor
Handles
Drip
chamber
Flanges
Pull back handles
1 . Remove the Flow Sensor Interface Cover and store for future use .
2 . Plug the flow sensor into the flow sensor interface located on the top rear part of the Pump .
3 . Attach the IVAC® Model 180 Flow Sensor to the drip chamber of the infusion set, by pulling back the handles . Refer to the illustration
above .
4 . Proceed with load, priming, and set-up instructions as described in this Directions For Use .
NOTE:
Ensure drip chamber is half full and upright .
A
Always attach the flow sensor before starting an infusion.
Avoid using the flow sensor in direct sunlight.
Always ensure lens is clean.
A
Always replace the Flow Sensor Interface Cover when the flow sensor is
disconnected.
Flow
Sensor
Interface
Cover