CORRECTIVE MAINTENANCE
Syringe Module
5A-10
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
5.2.8
Removing Internal Frame Assembly
1. Remove three (
3
) screws attaching
Internal Frame to Front Case Assembly.
2. Remove
IUI
Bracket Strap.
3. Remove one (
1
) screw attaching
IUI
Support Bracket to
IUI
Bracket.
4. Remove
IUI
Support Bracket and
IUI
Bracket.
5. Remove two (
2
) screws attaching
Internal Frame to Drivetrain Assembly.
6. Remove Internal Frame.
During Reassembly:
Route Linear Sensor Harness between
IUI
Bracket and Internal Frame.
5.2
DISASSEMBLY/REASSEMBLY
(Continued)
Drivetrain Assembly
IUI Bracket
Internal Frame
IUI Bracket
Strap
IUI Support Bracket
Front Case Assembly
Linear Sensor Harness
To replace an Internal Frame whose Barrel
Clamp Assembly has a park position “pin”
(rather than a “collar”), replace with Front
Case Assembly Kit (Table
9
).
CAUTION
-- Barrel Clamp Assembly
(Partial View)
--
Park Position “Collar”
Park Position “Pin”
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......