2.4.10 Priming
Enabling
Priming mode
allows a limited
volume of fluid to be delivered to prime the
administration set before connecting to a
patient or following a syringe change. When
priming, a single continuous press of the
PRIME
soft key delivers up to
2
mL of
priming fluid.
1. After accessing
System Config -
Syringe
options display, page
2
, press
Priming
soft key.
2. To change option setting, press soft key
next to applicable option (
Enable
or
Disable
).
• Option setting changes and
System
Config - Syringe
page
2
displays.
2.5
CONFIGURATION SETUP - SHARED
INFUSION - PUMP AND SYRINGE
MODULES
Refer to the
8000
/
8100
Series Technical
Service Manual.
2.6
CONFIGURATION OPTIONS AND
DEFAULTS - PCA MODULE
NOTE:
With the Profiles feature enabled, the
settings are configured independently for
each profile. A hospital-defined, best-
practice data set must be uploaded to enable
the Profiles feature. Date and Time is a
system setting and is the same in all profiles.
System Settings:
Refer to the PC Unit
section of the Alaris
®
System
DFU
.
PCA Module Settings:
Refer to the PCA
Module section of the Alaris
®
System
DFU
.
PCA Module Configuration Parameters:
Select system configuration settings for the
PCA Module during data set development,
then upload them as part of the data.
CHECKOUT AND CONFIGURATION
2-9
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
System Config - Pump 2 of 3
S
y
stem Configuration - S
y
ringe
Priming mode
Enable
Disable
Allows use of the
priming feature.
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
(Continued)
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......