2.3
CONFIGURATION OPTIONS AND
DEFAULTS - SYRINGE MODULE
NOTES:
• Changes to the factory default values are
retained after a power cycle.
• If
Factory Default
is
Yes
, then all the
configuration settings are set to their
factory default.
• If
Factory Default
is
No
, then one or
more of the configuration settings has
been changed. Setting
Factory Default
to
Yes
sets all configuration settings to
their factory default.
• With the Profiles feature enabled, the
settings are configured independently for
each profile. A hospital-defined, best-
practice data set must be uploaded to
enable the Profiles feature. Date and
Time is a system setting and is the same
in all profiles.
Refer to the Alaris
®
System
DFU
for the
following:
• System Settings
• Shared Infusion Settings (Pump Module
and Syringe Module)
• Syringe Module Settings
2.3.1
Factory Default Setting
To allow changes to System Configuration
parameters:
1. Hold
OPTIONS
key at power up.
2. Press
Factory default
soft key.
3. Press
No
soft key.
4. To accept change, press
EXIT
soft key.
• Option setting changes and main
System Config - Module
screen
displays.
2.4
CONFIGURATION SETUP - SYRINGE
MODULE
NOTES:
• Pressing the
EXIT
soft key while in a
System Config - Module
screen
immediately powers the system down, with
no “Powering Down” display.
• Pressing the
EXIT
soft key while in a
System Config - Syringe
screen returns
the display to the main
System Config -
Module
screen.
CHECKOUT AND CONFIGURATION
2-2
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
>Select an Option
or EXIT
EXIT
Factory default:
Yes
System Config - Module 1 of 2
PC Unit
Shared Infusion Settings
Pump Module
SPO2 Module
PAGE
DOWN
EXIT
S
y
stem Configuration
Factor
y
Default
Settings?
Yes
No
Selection of Yes sets all
S
y
stem Configuration
parameters to the factor
y
default setting.
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......