Table 6-3.
Error Code Matrix
Refer to the PC Unit / Pump Module (
8000
/
8100
Series) Technical Service Manual for the
Alaris
®
System error code matrix.
6600
Motor Watchdog Test Failure
6610
Motor Watchdog Failure
6620
Syringe Size Sensor Failure
6630
Plunger Force Not Tracking Pressure
6640
Plunger Force Sensor Failure
6650
Plunger Position Sensor Failure
6660
Plunger Position Failure
6670
Pressure Disc Sensor Failure (Syringe Module only)
6680
Timebases Out of Sync
6690
Pressure Sensor Current Monitor Failure
6700
Force Sensor Circuit Test Failure
6710
Digital Sensors Illegal State
6720
Motor Step Rate Failure
6730
Motor Step Sample Too Many Steps
6740
Nut Disengaged
6750
Nut Engage Failure
6760
Not Calibrated
6770
Pressure Sensor Circuit Test Failure
6780 PCA handset stuck (PCA Module only)
Table 6-2.
Subsystem Codes
Refer to the PC Unit / Pump Module (
8000
/
8100
Series) Technical Service Manual for Alaris
®
System
codes.
350 Safety System
351 Motor
352 Plunger Force
353 Plunger Position
354 Patient Pressure
355 Syringe Size
356 Digital Sensor
357 Keyboard
358 Keyboard Comm
TROUBLESHOOTING
6-6
Alaris
®
Syringe Module (8110 Series)
Alaris
®
PCA Module (8120 Series)
Technical Service Manual
Summary of Contents for Alaris 8110 series
Page 9: ...1 GENERAL INFORMATION...
Page 10: ......
Page 15: ...2 CHECKOUT CONFIGURATION...
Page 16: ......
Page 27: ...3 PREVENTIVE MAINTENANCE...
Page 28: ......
Page 31: ...4 PRINCIPLES OF OPERATION...
Page 32: ......
Page 41: ...5A CORRECTIVE MAINTENANCE SYRINGE MODULE...
Page 42: ......
Page 71: ...5B CORRECTIVE MAINTENANCE PCA MODULE...
Page 72: ......
Page 105: ...6 TROUBLESHOOTING...
Page 106: ......
Page 115: ...7A ILLUSTRATED PARTS BREAKDOWN SYRINGE MODULE...
Page 116: ......
Page 143: ...7B ILLUSTRATED PARTS BREAKDOWN PCA MODULE...
Page 144: ......
Page 171: ......