Alaris® Enteral Syringe 

Pump

Model: 8002ENT01

Directions For Use

en

Summary of Contents for 8002ENT01

Page 1: ...s Alaris Enteral Syringe Pump Model 8002ENT01 Directions For Use en...

Page 2: ...Features 6 Operating Precautions 7 Getting Started 9 Syringe Loading 11 Starting the Alaris Enteral Syringe Pump 14 Basic Features 15 Alarms and Warnings 18 Configured Options 20 Specifications 22 Rec...

Page 3: ...ut prior notification Indications The Alaris Enteral Syringe Pump is specifically indicated for the delivery of enteral therapies via Nasogastric Orogastric or Gastrostomy e g PEG Percutaneous Endosco...

Page 4: ...softkeys Positive Plunger Grippers Display Syringe Clamp ON OFF RUN HOLD PURGE MUTE PRESSURE OPTION Hook Finger Grips Hook Release lever for Rotating Cam Carrying Handle IR Communications port RS232 C...

Page 5: ...must not be connected to the patient Volume Delivered is not added to total volume delivered BOLUS fluid or drug delivered at an accelerated rate Alaris Enteral Syringe Pump is delivering Administrat...

Page 6: ...irect sprays of water up to 15 from vertical and protected against solid objects greater than 2 5mm Note IP33 applies if mains retainer kit part number 1000SP01294 is fitted r Alternating Current s De...

Page 7: ...Status Syringe type fitted Pressure Information Delivery Rate Volume Delivered Volume Delivered Option VTBD Option Screen Icons Symbol Description l Time remaining display icon Indicates time before s...

Page 8: ...ral wards neonatal wards paediatric wards critical and intensive care operating rooms accident and emergency rooms Ensure that the Alaris Enteral Syringe Pump is appropriately attached using the provi...

Page 9: ...e Pump in the vicinity of any Therapeutic Radiation Equipment Levels of radiation generated by the radiation therapy equipment such as Linear Accelerator may severely affect functioning of the Alaris...

Page 10: ...inge Pump to the AC power supply for at least 2 hours to ensure that the internal battery is charged verify that the S is lit Language Selection 1 On initial start up the Alaris Enteral Syringe Pump w...

Page 11: ...nsure the pole clamp is folded away and stored within the recess at the rear of the Alaris Enteral Syringe Pump before connecting to a Docking Station Workstation or when not in use Never mount the Al...

Page 12: ...necessary to deliver the fluid or medication w Warning Purge the Alaris Enteral Syringe Pump system before starting a delivery or after replacing a near empty syringe with a replacement syringe When P...

Page 13: ...dversely affect the accuracy of the delivery rate and may also affect Alaris Enteral Syringe Pump performance When drawing fluid into the syringe draw enough to compensate for any dead space volume in...

Page 14: ...ntil it reaches the plunger end 6 Release the finger grips Ensure that the plunger grippers are securing the plunger in place and the finger grip returns to its original position 7 Ensure that the syr...

Page 15: ...ep previous setup and go to step 8 Selecting YES will clear previous setup and will go to step 4 w Clear setup screen will only be displayed if a previous setup was used 4 Load the syringe according t...

Page 16: ...olus and continue delivery at the set rate Press the h button to stop the bolus delivery and place the pump on hold 6 If the bolus volume reaches the set bolus volume the bolus will stop and the pump...

Page 17: ...very will be allowed unless a new VTBD is set or current VTBD is cleared Clear Volume This option enables the volume delivered to be cleared 1 Press the VOLUME softkey to display the CLEAR VOLUME opti...

Page 18: ...on will be displayed SN The serial number of the Alaris Enteral Syringe Pump S W Software version of the Alaris Enteral Syringe Pump Adjust Alarm Volume To change the alarm volume if enabled 1 Press t...

Page 19: ...ct with the plunger button If there is no identifiable cause for the CHECK SYRINGE alarm s then the pump should be removed from clinical use and examined by Qualified Service Personnel in accordance w...

Page 20: ...dium Amber Three beeps will sound if the Alaris Enteral Syringe Pump has been left on for more than 2 minutes referred to as CALLBACK in the log without starting the operation Press the c button to si...

Page 21: ...ELECT softkey to return to the Configured Options menu Contrast This option is used to set the contrast on the Alaris Enteral Syringe Pump display 1 Select CONTRAST from the Configured Options menu us...

Page 22: ...set to be Hands On or Hands Free Pressure Display Displays the Pressure Icon on the main display Bolus Rate Default The default value for bolus rate 10ml h Bolus Rate Max The maximum value for bolus...

Page 23: ...ring BOLUS the pressure limit alarms are temporarily increased to their maximum level Bolus Volume Accuracy Bolus Volume Typical Typical Maximum Typical Minimum Pump Specification 0 1ml 1 9 6 2 7 3 10...

Page 24: ...ower Mean Time To Battery Empty from fully charged at 5ml h 23 C under normal conditions is 6 hours 95 lower confidence interval of 5 hours 50 minutes Charging takes 2 hours from discharge to 90 charg...

Page 25: ...601 2 24 Potential Equalisation Conductor The function of the Potential Equalisation Connector Conductor is to provide a direct connection between the pump and the potential equalisation busbar of the...

Page 26: ...Medicina Enteral PE05 PE10 PE20 PE60 Nutricair SE10 SE20 SE50 Terumo Enteral SS 30EO SS 50EO C 1 Please contact your local CareFusion representative for availability w CareFusion has characterized a...

Page 27: ...aximum Protection Against Electrical Shock Class 1 Classification Continuous Operation Supply to pump 115 230V 50 60Hz 60VA The Alaris DS Docking Station Product SKU 80283UNS00 xx Supply Voltage 230VA...

Page 28: ...its of measurement used in the calibration procedure are standard SI units The International System of Units Battery Operation The internal rechargeable battery allows continued operation when the AC...

Page 29: ...e oxide sterilise or immerse this Alaris Enteral Syringe Pump in any fluid If the Alaris Enteral Syringe Pump has visible cracks or damage to the case do not clean and immediately take out of service...

Page 30: ...ol the Alaris Enteral Syringe Pump using the RS232 interface at some distance from the Alaris Enteral Syringe Pump and hence remote from the patient responsibility for the control of the Alaris Entera...

Page 31: ...e Active Low 7V to 12V powers up the isolated RS232 circuitry Active High 7V to 12V Inactive Floating open circuit allows isolated RS232 circuitry to power down Isolation Socket Alaris Enteral Syringe...

Page 32: ...t of an occlusion when the Penta Enteral 60 ml feeding syringe is selected with a Pentaferte standard administration set Time to alarm 1 0 ml h Time to alarm 5 0 ml h hr min sec typical hr min sec typ...

Page 33: ...e trumpet Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered Short term fluctuations in rate accuracy may have clinical impact d...

Page 34: ...e Technical Service Manual 1000SM00001 is now available in electronic format on the World Wide Web at www carefusion co uk alaris technical A username and password are required to access our manuals P...

Page 35: ...3994 DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 902 555 6...

Page 36: ...ne of its subsidiaries and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction disclosure or use other than for...