background image

 

25 

 

Appendix D Recording Techniques 

                   

(Cont.)

 

 
 

b. By noting what happens to a sound or murmur when the bell is first applied lightly, and  
then with pressure, the examiner can judge, to a certain degree, the pitch of a sound or murmur. 

 

c. The high-pitched systolic murmur of mitral regurgitation is less likely to be affected by  pressure             
than medium-pitched, innocent systolic murmurs. 

 

d. With pressure, the faint, high-pitched, systolic murmur of mitral regurgitation is less affected 
than the first heart sound, and the masking effect of the first heart sound is thus diminished.  
The same is true of the high-pitched, early diastolic murmur of aortic regurgitation and 
accentuated second heart sound. 

 

e. In some very noisy hearts with both systolic and diastolic (because of the amount of  
sound in systole), it is difficult to be sure if there is a low-pitched diastolic murmur.   
With pressure, the total sound is diminished and the low-pitched murmur diastole may  
disappear. By holding the bell alternately lightly and heavily and by concentrating on diastole,  
the examiner can detect the diastolic rumble. 

Summary of Contents for SimulScope

Page 1: ...n System Cardionics 910 Bay Star Blvd Webster Texas 77598 USA Telephone 281 488 5901 Fax 281 488 3195 Email info cardionics com Web site www cardionics com Copyright 2012 Cardionics All rights reserve...

Page 2: ...a 3 wire AC power system containing a separate ground wire The ground wire must not be removed or defeated WARNING SHOCK HAZARD Do not remove protective covers on any components of the SimulScope Sys...

Page 3: ...e as a Classroom Infrared Emitter with CardioSim 10 3 0 USING THE SIMULSCOPE FOR TEACHING 15 3 1 Using the HeartMan Infrared Headphone 12 3 2 FOR BEDSIDE TEACHING 15 3 3 FOR CLASSROOM TEACHING 16 3 3...

Page 4: ...ith SimulScope The Heartman Infrared Headphone Figure 2 senses the infrared signal and converts it into an audible sound Each headphone has its own volume control and automatically shuts off when remo...

Page 5: ...smits the sounds as infrared energy to each HeartMan Infrared Headphone Figure 2 Heartman Infrared Headphone Figure 3 Front panel SimulScope Infrared Diodes A system of adjustable filters Figure 4 5 b...

Page 6: ...Figure 5 Graphic Equalizer Detail The above settings are considered normal for heart sounds The X settings are for normal heart sounds and the 0 settings for breath sounds Move the filters upward to...

Page 7: ...tman Infrared Headphones are also battery operated standard or alkaline 9V and will operate continuously for 150 or more hours at which time the battery should be replaced A low battery indicator flas...

Page 8: ...SimulScope is equipped with a Specialist Diaphragm stethoscope A Specialist Pediatric Bell cat No 360 7755 is available as an option Attach the stethoscope into the receptacle on the INPUT section of...

Page 9: ...CG leads to the snap on each of the three electrodes The end of the electrode lead is labeled with proper placement on the patient Connect each of the ECG leads to the connector end of the supplied pa...

Page 10: ...lessly connect the SimulScope to a PC This allows the user to not only send heart and lung sounds from the patient or SAM the Student Auscultation Manikin but also to a PC No wires or cables are neede...

Page 11: ...g and playback of Sounds and ECG Set Up for Recording To record and playback both heart sounds and ECG you will need an Audio Interface Module UA 1ex see figure 11 As shown in the below diagram connec...

Page 12: ...Select Number of Channels Select two 2 channels Set up for Play back Refer to the Playback section of Figure 11 Connect the Audio Interface Module to the Stethoscope input on the rear panel of the Sim...

Page 13: ...rared Emitter with the CardioSim Digital Heart Sound Teaching System To use the SimulScope as a classroom infrared emitter with the CardioSim Digital Heart Sound Teaching System connect the Stethoscop...

Page 14: ...use the CardioSim in the normal manner Turn the CardioSim speaker volume OFF and listen using Heartman Infrared Headphones Use Cardionics cable Cat No 711 7780 not supplied Connect Stethoscope IN to C...

Page 15: ...igital Heart Sound System or with SAM the Student Auscultation Manikin At the bedside attach the stethoscope to the Stethoscope INPUT on the rear panel then turn the volume control ON We recommend tha...

Page 16: ...ssful use The SimulScope should be aimed at the students The instructor can stand behind the SimulScope if desired If for any reason the reception is not sufficient try aiming the SimulScope at a mirr...

Page 17: ...M the Student Auscultation Manikin following instructions in the SAM Operator s manual Place the stethoscope of the SimulScope on SAM as you would a live patient You may listen to heart sounds at the...

Page 18: ...tput of Harvey located on the lower right side of the Harvey cabinet to the Stethoscope Input of the SimulScope located on the rear panel of the SimulScope Use Cardionics cable Cat No 711 0254 Turn ON...

Page 19: ...tions a sending station or location and a receiving station 4 1 Using the SimulScope with Polycom or Tandberg Teleconferencing Systems For sending connect the output of the SimulScope to the audio inp...

Page 20: ...SimulScope may be attached to the IR output of the CardioSim and used as a classroom infrared emitter With the PneumoSim Digital Breath Sound Teaching System The SimulScope will transmit sounds produc...

Page 21: ...ethoscope with Specialist stainless steel bell diaphragm Cables 711 0243 SimulScope to PC Mic Input for recording or display 711 0244 PC to SimulScope Stethoscope IN for playback 711 0246 Polycom or T...

Page 22: ...ergy and cause weak reception if direct contact between the headphone is not maintained Pointing the face of the SimulScope toward a mirror will increase reception Abnormal White Noise See Weak Recept...

Page 23: ...There are several possible solutions as follows 1 turn off the lights 2 move the SimulScope transmitter so that it is not located directly under a florescent light Ask the listeners not to stand dire...

Page 24: ...who may have a friction rub This position may be awkward when the patient is in bed or on the examining table and supporting himself herself on the elbows It is much more satisfactory to have the pati...

Page 25: ...urgitation is less affected than the first heart sound and the masking effect of the first heart sound is thus diminished The same is true of the high pitched early diastolic murmur of aortic regurgit...

Page 26: ...tection AUDIO FREQUENCY RESPONSE 30 Hz TO 4 KHz 3 dB with equalizer controls centered AUDIO EQUALIZER Provides 12 dB of gain adjustment via individual slide control for each band Band center frequenci...

Page 27: ...s orange INDICATOR to indicate low battery condition with approximately 30 minutes operating time remaining AC POWER SUPPLY Wall mount 120VAC power supply allows unit to operate BATTERY CHARGER while...

Page 28: ...limitations on how long an implied warranty lasts so the above limitations may not apply to you 4 This warranty is limited to repair of the product or replacement thereof at the discretion of Cardioni...

Page 29: ...service Ship package prepaid to Cardionics 910 Bay Star Blvd Webster Texas 77598 USA We will endeavor to complete repairs within fifteen 15 working days from date of receipt of product by Cardionics O...

Page 30: ...pply 4 Figure 7 Rear panel view 5 Figure 8 Connecting ECG cable 6 Figure 9 StethView Digital Recording System 7 Figure 10 Using the HeartMan Wireless PC Receiver 8 Figure 11 Recording and Playback of...

Page 31: ...I Index 28 Infrared diodes 2 L List of Figures 27 P Parts list 18 Pin outs rear panel 24 Playback Computer 8 PneumoSim digital lung sound simulator Using SimulScope with 17 Product description 1 R Re...

Reviews: