8400-M022-O1 Rev K
Model 190 Indicator
64
Peak Hold Function
With the Peak Hold function enabled during setup of the indicator, the function must be
activated before it can be used. Refer to the section Function Setup, the
FUnC
(
FUnC?
)
parameter to enable the
Fn/
key Hold function.
NOTE:
Only one of the functions Hold, Count, Peak Hold, Checkweigher, or Live
Weight can be active at a time.
Activate Hold Function
1.
Press and hold the
Fn/
key until the display blanks.
2.
Once the display blanks, release the
Fn/
key
3.
The display will change to show
PHold
X, the prompt, and status of the Peak Hold
function.
4.
If the setting displayed is acceptable, press the
TARE
key to save it. Otherwise, use
the
Fn/
key to toggle to a new setting and press the
TARE
key to save it.
1
= Peak Hold Function Active
0
= Peak Hold Function Not Active
5.
Repeatedly press the
TARE
key to step through the remaining function prompts until
the indicator returns to the weight display.
Peak Hold Function Operation
When the Peak Hold function is active, the largest weight value measured since Peak Hold
was activated (or cleared) will be stored.
1.
To display the peak value, momentarily press the
Fn/
key.
2.
To zero the peak value, make sure there is no load on the scale and press the
ZERO
key.
3.
Press the
NET/GROSS
key to return to the weight display.