8400-M022-O1 Rev K
Model 190 Indicator
119
Setup
Enter
SETUP
and proceed to the Options Setup.
option
With
option
displayed, press the
TARE
key. The display will change to show
opt=
.
Proceed to the
opt=
parameter.
opt?
With
opt?
displayed, press the
TARE
key. The display will change to
no
. Press the
Fn/
key to toggle to
yes
and press the
TARE
key. The display will change to
opt=
.
Proceed to the
opt=
parameter.
Otherwise, to skip the Options Setup, press the
TARE
key to advance to the
fUnC?
.
prompt.
oPt=
(Option)
Press the
TARE
key to show the current setting. If the setting displayed is acceptable,
press the
TARE
key again to save it.
Otherwise, use the
Fn/▲
key to toggle to a new setting and press the
TARE
key to save
it. Allowable values are:
0 = No Option Card Installed
1 = Serial Interface, Ethernet, Wi-Fi, or USB Option Card Installed
2 = not applicable
3 = not applicable
4 = Digital to Analog Converter Option Card Installed
rAnGE=
Press the
TARE
key to show the current setting. If the setting displayed is acceptable,
press the
TARE
key again to save it.
Otherwise, use the
Fn/▲
key to toggle to a new setting and press the
TARE
key to save
it. Allowable values are:
0 = 4-20 ma
1 = 0-20 ma
2 = 0-24 ma
3 = 0-10 volts
4 = 0-5 volts
5 = -10 to +10 volts
6 = -5 to +5 volts