IVAC® Volumetric Pump (Models 571 & 572)
81/104
1000SM00018 Issue 2
Spare Parts Replacement Procedures
Pole Clamp Assembly
Replacement Procedure
1. Referring to the 'Mounting steps' 1 and 2 on the next page, unlock the catch, levering it open with a blunt object such
as a coin if necessary.
2. Push the adapter bracket away from the backstop to release it then lift the pole clamp adapter up and off the screw to
remove it.
3. Using a hex key, loosen the one fixing screw at the bottom centre of the pump. Remove the screw.
4. Reassemble in reverse order.
Spare Parts
Item
Description
Part Number
A
POLE ADAPTER
6767
B
TRANSPORT ADAPTOR WC
5767C
C
IV POLE CLAMP ACCESORY
BC100A
D
POLE CLAMP HEX SCREW 56X/57X
SCR-1347
Catch
(
D
) Screw
(
B
) Transport Adapter
(
D
) Screw
Catch
(
A
) Pole Adapter
(
C
) Clamp
Note:
Transport Adapter
(
B
) is used in
conjunction with Clamp (
C
)
Refitting notes:
1) Do not over-tighten the one fixing
screw when securing snugly with hex
key.
2) See also 'Mounting steps' on the next
page.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...