IVAC® Volumetric Pump (Models 571 & 572)
17/104
1000SM00018 Issue 2
Configuration & Calibration
Calibration Procedures
Equipment required:
50ml syringe
Pressure gauge
"52" Series IV infusion set
Procedure:
1. Remove the battery cover from the pump to
allow access to the Accessory Interface Board.
See Chapter 6, 'Spare Parts Procedures' for
instructions.
2. Prepare a modified IV infusion set as follows:
Completely remove the clear membrane
cover from the white pressure sensing disc.
Ensure no membrane remains as this could
compromise the integrity of the vacuum
Attach the air-filled IV infusion set to the
syringe. Use the luer fitting on the distal end
of the set or cut approximately 1 inch (2.5
cm) off the distal end and press the tubing
firmly into the tip of the syringe. Ensure a
good seal is formed between the two devices
Splice a T-fitting into the IV infusion set
tubing approximately 1 foot (30 cm) up from
the syringe attachment. Connect the pressure
gauge to the T-fitting port
Push the syringe plunger in until it stops
3. Load the modified IV infusion set into the pump
and close the door. Ensure that the set tubing is
not kinked.
4. Enter self-test mode and select test
027
. See 'Entering
Self-Test Mode' for instructions.
5. Press the
RUN/HOLD
switch to enter the Vacuum Display mode and display
VAC MMHG
.
6. Pull the syringe plunger out until the pressure gauge reads
-95
mmHg and immediately clamp off tubing below the
pump to hold the vacuum stable.
7. Adjust the vacuum detect potentiometer
R45
on the Accessory Interface Board until the numeric display on the
pump reads
-95
± 5 mmHg
.
Note:
R45
is the upper potentiometer on the Accessory Interface Board.
8. Perform the 'Vacuum Retention Verification' procedure as detailed in
Chapter 3, 'Routine Maintenance'.
Vacuum System Calibration (Test 27)
Syringe
Pressure
Gauge
Proximal
End of Set
Selecting Pressure Units
The pressure units can be configured to be displayed as millimetres of mercury (mmHg) or
as centimeters of water (cmH20).
To select the pressure units:
Set the Pressure Unit Selector switch on the rear panel to the required position.
Pressure Unit
Selector Switch
Do
not
adjust the lower
potentiometer, R4, this
is pre-adjusted during
manufacture.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...