IVAC® Volumetric Pump (Models 571 & 572)
12/104
1000SM00018 Issue 2
Introduction and Start Up
Verifying a value in one mode while operating in another
1. Press and release the appropriate (current) mode switch. For example,
SEC
.
2. Within two seconds press the switch representing the information to be displayed. For example,
PRI
.
Changing the infusion manually to primary mode
1. Press the
RUN/HOLD
switch to set pump on hold.
2. Press the
PRI
switch; pump sounds a three beep verification tone and goes into PRIMARY mode.
Secondary Infusion (Model 572) continued
Additional Programming during Secondary Infusion
Flow Sensor
A flow sensor (part number 180) is used to detect an empty solution container. It is connected to the drip chamber and
when a drop is detected in the drip chamber
an illuminated ("0") appears in the Information Display
.
When not in use, the flow sensor can be stored on the pump handle.
Pole Clamp Accessories
The following pole clamp accessories allow the pump to be mounted on conventional IV poles:
Universal Pole Adapter (part number 6767)
Pole Adapter (part number 5767A)*
Pole Clamp Accessory (part number BC100A)*
* must be used in conjunction with each other.
Refer to Chapter 6, 'Spare Parts Replacement Procedures' for assembly information.
Summary of Contents for IVAC 571
Page 1: ...IVAC Variable Pressure Volumetric Pump Models 571 572 Technical Service Manual...
Page 37: ...Chapter 4 Troubleshooting In this chapter Introduction 38 Error Messages 38 General Faults 42...
Page 97: ...Appendix C Configured Options...
Page 99: ...Appendix D Disposal...
Page 101: ...Appendix E Service Contacts...
Page 103: ...Appendix F Document History...