4. Press the soft key to select the
Limit
Mode Setup
display, or press
CANCEL
to exit the
System Config - EtCO
2
display without changing the value.
2.4.3
Waveform Height
1. To change the waveform height, you can
toggle its value from the
System Config
- EtCO
2
display, or you can use the
Waveform height
soft key.
2.
To toggle the value:
From the
System
Config - EtCO
2
display, press the soft
key for the waveform height value.
Every key press toggles the value
between 60 and 99 mmHg.
3.
To use the soft key:
From the
System
Config - EtCO
2
display, press the
Waveform height
soft key.
4. Press the soft key to select the
waveform height, or press
CANCEL
to
exit the
Waveform Height
display
without changing the value.
2.4.4
Waveform Time Scale
1. To change the waveform time scale, you
can toggle its value from the
System
Config - EtCO
2
display, or you can use
the
Waveform time scale
soft key.
CHECKOUT AND CONFIGURATION
2-5
Alaris
®
EtCO
2
Module (8300 Series)
Technical Service Manual
2.4
CONFIGURATION SETUP
(Continued)
2.4.2
Limit Mode
(Continued)
Press the
Waveform height
soft key to toggle
between the two
acceptable values
Press the
Waveform time
scale
soft key to
toggle between the
two acceptable
values
Summary of Contents for Alaris 8300 Series
Page 7: ...1 GENERAL INFORMATION...
Page 8: ......
Page 11: ...2 CHECKOUT CONFIGURATION...
Page 12: ......
Page 21: ...3 PREVENTIVE MAINTENANCE...
Page 22: ......
Page 25: ...4 PRINCIPLES OF OPERATION...
Page 26: ......
Page 29: ...5 CORRECTIVE MAINTENANCE...
Page 30: ......
Page 47: ...6 TROUBLESHOOTING...
Page 48: ......
Page 53: ...7 ILLUSTRATED PARTS BREAKDOWN...
Page 54: ......
Page 67: ......