5-10
Programming
Alaris
®
System Directions for Use
EtCO
2
Module Section
Changing Waveform Height
1.
Press
Waveform height
soft key.
A
Waveform height: 80 mmHg
Waveform time scale: 5 sec
Limit Mode:
Adult
Channel Options
>Select an Option or
EXIT
EXIT
2.
To change
Waveform Height
, select applicable range
limit.
•
60
mmHg:
Displays a waveform for
E
t
CO
2
values
within
0 – 60
mmHg range. If
E
t
CO
2
value exceeds that
range,
Waveform Out of Range; Adjust Scaling
message displays until waveform falls back into range
or
0 – 99
mmHg option is selected.
•
99
mmHg:
Displays a waveform for full
E
t
CO
2
value
range,
0 – 99
mmHg.
3.
To return to
ETCO2
Main
display, press
EXIT
soft key.
A
A
60 mmHg
Waveform Height
EXIT
>Select an Option or
EXIT
99 mmHg
Channel Options
(Continued)
2.
To change
Limit Mode Setup
, press applicable soft key.
OR
To leave
Limit Mode Setup
unchanged and return to
ETCO2
Main
display, press
EXIT
soft key.
A
A
Adult
Neonatal
Limit Mode Setup
EXIT
>Select an Option or
EXIT
Changing Limit Mode
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......