Programming
4-7
Alaris
®
System Directions for Use
SpO
2
Modules Section
Monitoring Mode
(Continued)
2.
To change a limit setting, press soft key next to applicable
parameter.
A
%SPO2
HIGH
%SPO2
LOW
Off
90
150
50
>Select Parameter Limit
PULSE
HIGH
PULSE
LOW
CONFIRM
Adult Alarm Limits
Setting Alarm Limits
(Continued)
3.
Enter a numeric value for selected alarm limit.
•
%SPO2 HIGH
limit can be
Off
or a numeric value.
4.
To move to next limit, press
ENTER
key.
5.
To confirm alarm settings and return to
SPO2
Main
display, press
CONFIRM
soft key.
6.
To return to Main Display, press
MAIN SCREEN
soft key.
A
A
%SPO2
HIGH
%SPO2
LOW
>Press CONFIRM to
Apply Changes
_98
90
150
50
Off
PULSE
HIGH
PULSE
LOW
Adult Alarm Limits
CONFIRM
Navigating Trend Data
1.
To view
Trend Data
, press
TREND
soft key.
A
Adult
Mode
82
% SPO2
97
Off
150
90
50
LIMITS
TREND
MAIN
SCREEN
PULSE
RATE
SPO2 Main
SatSeconds (% to Alarm)
0
100
>Press ENTER to Retain
Current View
2.
To navigate from page to page, press
PAGE UP
and
PAGE
DOWN
soft keys.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......