3-60
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Alarms, Errors, Messages
(Continued)
Messages
Errors
Channel Error
Syringe Calibration Required
Syringe Driver Head Error
Error detected. Operation stops on
affected module.
Error on infusing module indicating
calibration is required. Infusion
stops on affected module.
CALIBRATE
scrolls in Message
Display.
Noninfusing module, with plunger
grippers open, senses excessive
pressure being applied downward
on Drive Head.
OCCLUSION
scrolls
in Message Display.
To silence alarm and continue
operation of unaffected modules,
press
CONFIRM
soft key. Replace
module with an operational
instrument, as required. Service by
qualified personnel is required.
To silence alarm and continue
operation of unaffected modules,
press
CONFIRM
soft key. Replace
module with an operational
instrument, as required. Service by
qualified personnel is required.
To silence alarm and continue
normal operation, press
CONFIRM
soft key.
Error
Meaning
Response
Message
Meaning
Response
Bolus Complete
Infusion Complete
Load Complete
Max Limit Reached
Current bolus dose completed.
Channel running in continuous
dose if programmed.
Current infusion completed.
Current loading dose completed.
Infusion mode menu available or
programmed infusion running.
Programmed maximum limit has
been reached over time period
specified. Infusion paused until
time limit has expired.
None
Set up a new infusion or press
CHANNEL OFF
key.
None
To silence alarm, press
SILENCE
key.
To change
Max Limit
, press
CHANNEL SELECT
, press
SETUP
soft key, and unlock door or enter
Authorization Code applicable for
current Security Access Level.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......