General Information
General Information
3-39
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
• The PCA Module is designed to
stop fluid flow under
alarm conditions
. Periodic patient monitoring must be
performed to ensure the infusion is proceeding as
expected. It is a
positive displacement delivery system
,
capable of developing positive fluid pressures to overcome
widely varying resistances to flow encountered in practice,
including resistances to flow imposed by small gauge
catheters, filters and intra-arterial infusion. It is neither
designed nor intended to detect infiltrations and will not
alarm under infiltration conditions.
• The use of positive displacement infusion devices ported
together with
gravity flow infusion
systems into a
common
IV
site may impede the flow of common “gravity
only” systems, affecting their performance. Hospital/facility
personnel must ensure the performance of the common
IV
site is satisfactory under these circumstances.
•
Each time the Alaris
®
System is turned on
, verify and/or
set the monitoring mode, resistance alert, and/or pressure
alarm limit. If the monitoring mode, resistance alert, and/or
pressure alarm limit are not verified, the instrument may
not operate within the desired occlusion detection
parameter(s).
WARNINGS
Warnings and Cautions
General
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......