Occlusion Pressure
A complete range of downstream occlusion detection options is
provided.
•
Pump mode:
Downstream occlusion alarm threshold is
525
mmHg at flow rates of
30
mL/h or greater. For rates
<30
mL/h, occlusion pressure is rate-dependent, to ensure rapid
response to occlusions.
•
Selectable pressure mode:
Downstream occlusion alarm
threshold can be adjusted in
25
mmHg increments, up to
maximum occlusion pressure of
525
mmHg.
•
Auto-Restart:
(See "Auto-Restart" definition.)
In addition, Alaris
®
System provides fluid-side occlusion detection.
Secondary Infusions
Dual rate sequential piggyback (secondary) infusions may be
infused, with limits, at delivery rates and volumes independent of
primary infusion parameters. Automatic changeover occurs to
primary infusion parameters when secondary infusion is complete if
an Alaris
®
Pump Module/Gemini Infusion System check valve
administration set is used.
2-84
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Pump Module
(Continued)
Features and Displays
(Continued)
Features and Definitions
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
All Mode
When
ALL
is selected as the volume to be infused (
VTBI
), entire
contents of the syringe will be delivered.
Auto Pressure
When enabled and a pressure sensing disc is in use, Auto
Pressure option is displayed in Pressure Limit screen. Auto
Pressure automatically sets alarm limit for a shorter time to alarm,
as follows:
• If current pressure is
100
mmHg or less, system adds
30
mmHg to current pressure, to create a new alarm limit.
• If current pressure is greater than
100
mmHg, system adds
30%
to current pressure, to create a new alarm limit.
Auto Pressure Limit Adjustment
When a bolus is delivered, pressure alarm limits are temporarily
raised to maximum limit.
Syringe Module
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......