Since, by definition, a multidose infusion will not be infusing
for a programmed period of time, it is assumed that another
infusing
IV
line will keep the vein open until the beginning of
the first dose and between subsequent doses. There is no
keep vein open (
KVO
) infusion at the completion of a
programmed
Delay until
infusion.
Syringe Module:
ALL
Mode is not supported in Multidose
Mode.
2-62
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Multidose Mode
• The Multidose feature is to be
used only by
personnel
properly trained
in using
multidose infusions.
•
Caution labels
, which clearly
differentiate single dose and
multidose containers, must be
utilized.
• Single dose piggybacking
systems employing
check
valve sets
are not designed
for use with multidose containers.
WARNINGS
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
3.
Press
CONFIRM
soft key.
•
PAUSE
scrolls in Message Display.
•
PAUSED
appears on Main Display.
• Yellow Standby Status Indicator illuminates.
• After
2
minutes:
PAUSE
-
RESTART CHANNEL
visual
and audio prompts begin, and yellow Standby Status
Indicator flashes.
4.
To reinitiate infusion:
• Press
RESTART
key.
OR
• Press
CHANNEL SELECT
key and then
START
soft key.
NOTES:
Using the
Pause
function in the Delay Options screen is the
same as pressing the
PAUSE
key on the Syringe Module.
The time displayed in the upper right corner of the screen is
the time of day in a
24
-hour clock format (military time).
Delay Options
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Pausing Infusion
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......