The following procedure should be used only when:
• drug to be infused is not listed in Drug Library,
• primary infusion is running, and
• a check valve administration set is being used.
To program a primary infusion, reference ”Primary Infusion -
With Guardrails
Suite MX Protection”, “IV Fluid Infusion”
procedure. To program a basic infusion, reference ”Infusion -
1.
Press
SECONDARY
soft key and then
BASIC SEC
soft key.
Continue with next step.
2.
Enter secondary infusion rate or duration, as follows:
• To enter secondary infusion rate, press
RATE
soft key
and use numeric data entry keys.
• To enter duration, press
DURATION
soft key and use
numeric data entry keys.
3.
To enter secondary volume to be infused, press
VTBI
soft
key and use numeric data entry keys.
4.
Open clamp on secondary administration set.
5.
Verify correct infusion parameters and press
START
soft
key.
Programming
2-51
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Secondary Infusion - NO
Guardrails
®
Suite MX Protection
(
Pump Module
)
Infusion
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
1.
Press
CHANNEL SELECT
key.
2.
Press
PRIMARY
soft key.
3.
To change primary infusion parameter, press applicable
soft key (
RATE
or
VTBI
), and use numeric data entry keys.
Changing Primary Infusion Parameter
Introduction and Setup
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......