70-00569-02 G
Powerheart® G5 Automated External Defibrillator
A-6
RescueCoach™ voice and text prompts
Analysis interrupted. Stop
patient motion.
ANALYSIS INTERRUPTED
STOP PATIENT MOTION
If the AED detects ECG noise
artefacts, stop moving or
touching the patient. Remove
other electronic devices from
the vicinity.
X
X
X
Table A-4: Analysis (continued)
Text display
Prompt level
Voice prompt
Line 1
Line 2
Situation
Adv
Std
Bas
Table A-5: Delivering therapy—G5 semi-automatic
Text display
Prompt level
Voice prompt
Line 1
Line 2
Situation
Adv
Std
Bas
Press the red flashing
button to deliver shock.
PRESS BUTTON
TO DELIVER SHOCK
Prompts after the AED is fully
charged and ready to deliver the
defibrillation shock. The red Shock
button flashes and the phrase
repeats for 30 seconds or until
the SHOCK button is pressed.
X
X
X
Shock delivered.
SHOCK DELIVERED
Prompts when the shock is
delivered.
X
X
X
Rhythm changed. Shock
cancelled.
RHYTHM CHANGED
SHOCK CANCELLED
Notifies the rescuer when the
AED detects a rhythm change
and cancels the shock.
X
X
X
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......