70-00569-02 G
Powerheart® G5 Automated External Defibrillator
5-6
Troubleshooting
Diagnostic mode messages
The Diagnostic mode provides details on the maintenance and service conditions of the AED.
For example, if the AED is not Rescue Ready, Diagnostic mode displays additional information
about the status.
To enter Diagnostic mode:
◆
Press and hold both buttons on the AED’s display
panel for three seconds.
The following prompts appear when the AED is in Diagnostic mode. Use the table to resolve the
reported conditions.
Press
Text display
Voice prompt
Line 1
Line 2
Situation
Resolution
Diagnostic mode.
DIAGNOSTIC MODE
The AED enters Diagnostic
mode.
Not applicable
SERVICE REQUIRED
CONTACT TECH SUPPORT
The AED detects a condition
that can prevent the AED from
continuing a rescue.
Contact Cardiac Science
Technical Support or your
local representative
immediately.
EXTREMELY LOW BATTERY
REPLACE BATTERY
The battery charge is too low
to support a rescue.
Replace the battery
immediately. If the battery
charge is completely depleted,
all AED activity ends.
Summary of Contents for Powerheart G5
Page 1: ...User Guide 70 00569 02 G Cardiac Science Powerheart G5 Automated External Defibrillator...
Page 2: ......
Page 3: ...USER GUIDE POWERHEART G5 AUTOMATED EXTERNAL DEFIBRILLATOR 70 00569 02 G...
Page 8: ...Contents 70 00569 02 G vi Contents...
Page 14: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 1 6 About the AED...
Page 22: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 2 8 Steps of a Rescue...
Page 36: ...70 00569 02 G Powerheart G5 Automated External Defibrillator 4 4 AED Features...
Page 93: ......
Page 94: ......
Page 95: ......